Instrument Function
R&S
®
SMBV100A
153
Operating Manual 1176.8016.02 ─ 17
Note
The character of the first key you pressed is immediately added in the input
field. Prior to inserting the password delete this entry.
●
In remote control mode, send the command
SYST:ULOC ENABled
to release all
locks at once.
Alternatively, you can use the command
SYST:KLOC OFF
to unlock the keyboard,
or
SYST:DLOC OFF
to release the display.
Via remote control, there is no password required.
Remote command:
Security Password
Enters the password that is required to enable or to disable the settings protected by a
security password. Default is '123456'.
Note:
It is highly recommended that you to change the default security password
before connecting the instrument to the network.
All settings are only accepted after the "Accept" button is pressed.
Accept
Applies the modified settings, provided the security password is entered correctly.
Note:
This action can not be undone. Keep in mind, that a changed password is never
reset, even if you perform a factory preset.
Bluetooth Pin
Sets the Bluetooth pin of an external Bluetooth device. The pin is required to enable
remote control via an external Bluetooth device.
Requires a Bluetooth adapter (recommended extra, see data sheet) .
The interface parameters are identical to the RS232 interface parameters and can be
set in the "Remote Channel Settings" dialog (see
5.2.3.19
Save/Recall
The "Save/Recall" submenu can also be called up with the FILE key and is therefore
described in the section of this key (see
Chapter 5.2.8, "Storing and Loading Instru-
General Instrument Settings