Remote Control Commands
R&S
®
SMBV100A
540
Operating Manual 1176.8016.02 ─ 17
Parameters:
<Ignore>
OFF | ZERO | ONE
OFF
Pattern Ignore is not active.
ONE
Bit sequences consisting of 32 or more consecutive "1" data are
not used (i.e. ignored) for the BER mesaurement.
ZERO
Bit sequence consisting of 32 or more consecutive "0" data are
not used (i.e. ignored) for the BER measurement.
*RST:
OFF
Example:
BERT:SET:IGN ONE
specifies that bit sequences of "1" data at least 32 bits long are
not used for the measurement.
Manual operation:
See
:BLER:SETup:MCOunt
<MCount>
:BERT:SETup:MCOunt
<MCount>
Enters the number of transmitted data bits/data blocks to be checked before the mea-
surement is terminated. With a BER measurement, data that was suppressed by
BERT:SETup:DENable
is not counted. This termination criterion always terminates
the measurement after the specified number of data bits/data blocks. Starting from this
point, the fourth value is output with 1 (= terminate measurement) if the result is quer-
ied with
:BERT|BLER:RES
. If the continuous measurement mode (
BERT|BLER:SEQ
AUTO
) has been selected, the measurement is restarted once the results have been
queried.
Parameters:
<MCount>
integer
Range:
0 to 4294967295
*RST:
1E7
Example:
BERT:SET:MCO 1E6
sets a measurement over 1000000 data bits.
Example:
BLER:SET:MCO 1E3
sets a measurement over 1000 data blocks.
Manual operation:
See
:BLER:SETup:MERRor
<MError>
:BERT:SETup:MERRor
<Merror>
Enters the number of errors to occur before the measurement is terminated. This termi-
nation criterion always terminates the measurement after the specified number of
errors. Starting from this point, the fourth value is output with 1 (= terminate measure-
ment) if the measurement result is queried with
:BERT|BLER:RES
.
BERT and BLER Subsystems