Remote Control Commands
R&S
®
SMBV100A
617
Operating Manual 1176.8016.02 ─ 17
Parameters:
<Offset>
float
Range:
-1 to 1
Increment: 0.001
*RST:
0
Default unit: FS
Example:
BB:ARB:TSIG:RECT:OFFS 0.5
sets the DC component
Manual operation:
See
"Offset DC (Rect Test Signal)"
[:SOURce<hw>]:BB:ARBitrary:TSIGnal:RECTangle:SAMPles
<Samples>
Sets the number of sample values required for the rectangular signal per period.
Parameters:
<Samples>
integer
Range:
4 to 1000
*RST:
100
Example:
BB:ARB:TSIG:RECT:SAMP 400
sets 400 samples per period
Manual operation:
See
"Samples per Period (Rect Test Signal)"
[:SOURce<hw>]:BB:ARBitrary:TSIGnal:SINE:FREQuency
<Frequency>
Sets the frequency of the simple sinusoidal test signal. This signal is used as output via
the I channel. A sine wave of the same frequency but optionally phase-shifted is gener-
ated on the Q path (
ARB:TSIGnal:SINE:POFFset
).
Parameters:
<Frequency>
float
Range:
100 to depends on hardware
Increment: 0.01
*RST:
1000
Default unit: Hz
Example:
BB:ARB:TSIG:SINE:FREQ 100 kHz
sets a sine signal of 100 kHz.
Manual operation:
See
"Frequency (Sine Test Signal)"
[:SOURce<hw>]:BB:ARBitrary:TSIGnal:SINE:PHASe
<Phase>
The command sets the phase offset of the sine wave on the Q channel relative to the
sine wave on the I channel.
SOURce Subsystem