Remote command:
[:SOURce]:INPut:CLOCk:IMPedance
Baseband Trigger Input Slope
Selects the polarity of the active slope of an externally applied trigger signal.
This setting affects the INST TRIG input connector (BNC connector at the rear of the
instrument).
Remote command:
[:SOURce]:INPut:TRIGger:BBANd:SLOPe
Ext. Trigger Input Slope
Sets the polarity of the active slope of an externally applied instrument trigger.
This setting affects the TRIG input (BNC connector at the rear of the instrument).
"Positive"
activates the rising edge of the trigger signal.
"Negative"
activates the falling edge of the trigger signal.
Remote command:
Clock Input Slope
Sets the polarity of the active slope of an externally applied symbol clock pulse.
This setting affects the CLOCK input (BNC connector).
Remote command:
Automatic Delay Compensation
While working in a master-slave mode, the instruments are connected as daisy chain
(see also
Chapter 5.9.2.4, "Synchronous Signal Generation"
tings for automatic delay compensation enable each of the slaves to calculate and
compensate the expected timing delay caused by the used cables. The calculation
assumes an ordinary BNC cables that have a dielectrics with a permittivity of
Ɛ
r = 2.7.
General Instrument Settings