Remote Control Commands
R&S
®
SMBV100A
721
Operating Manual 1176.8016.02 ─ 17
Usage:
Query only
Manual operation:
See
[:SOURce<hw>]:BBIN:SRATe[:ACTual]
<Actual>
Sets the sample rate of the external digital baseband signal.
Parameters:
<Actual>
float
Range:
400 to 200E6
Increment: 0.001
*RST:
100E6
Example:
:SOURce:BBIN:SRATe:ACTual 50E6
Manual operation:
See
[:SOURce<hw>]:BBIN:SRATe:FIFO[:STATus]?
This command indicates the state of the signal transmission.
Return values:
<Status>
OK | OFLow | URUN
OK
Resampling is working properly.
OFL = Overflow
The external sample rate frequency is too high which causes an
overflow in signal transmission.
URUN = Underrun
The external sample rate frequency is too low which causes an
underrun in signal transmission.
*RST:
OK
Example:
BBIN:SRAT:FIFO:STAT
queries the FIFO Status of the sample rate.
Usage:
Query only
Manual operation:
See
[:SOURce<hw>]:BBIN:SRATe:SOURce
<Source>
This command queries the source for estimating the sample rate of the digital input sig-
nal or defining it by the user.
Note
: With a connected
R&S EX-IQ-Box
either the internal clock source (User
defined) or an external clock (Digital I/Q In) can be selected in the settings dialog of the
R&S EX-IQ-Box. Therefore, the sample rate source is only indicated in this field.
SOURce Subsystem