Instrument Function
R&S
®
SMBV100A
456
Operating Manual 1176.8016.02 ─ 17
Remote command:
[:SOURce<hw>]:BBIN:DIGital:ASETting:STATe
Measurement Period
Sets the recording duration for measuring the baseband input signal by
"Auto
Level
Set". For accurate level measurements, the measurement period should
be set to a time value that is long enough to capture several periods of the input signal.
Remote command:
Auto Level Set
Starts measuring the input signal. The measurement estimates the crest factor, the
peak power and the rms power. The estimated values are automatically entered into
the input fields "Crest
Factor", "Peak Level" and "Level". Using these estimated values
the internal gain control adjusts the input signal gain to achieve an optimal dynamic
range.
See also
"Fulfilling the prerequisites for a correct signal processing of the externally
Remote command:
[:SOURce<hw>]:BBIN:ALEVel:EXECute
Crest Factor
Sets the crest factor of the external analog or digital baseband signal. Indicates the
crest factor acquired with "Auto Level Set".
See also
"Fulfilling the prerequisites for a correct signal processing of the externally
Remote command:
Peak Level
Enters the peak level of the external analog or digital baseband signal relative to full
scale of 0.5 V (in terms of dB full scale).
Indicates the peak level acquired with "Auto Level Set".
Remote command:
Level
Indicates the estimated rms level acquired with "Auto Level Set".
Remote command:
Signal Monitoring
The section "Signal Monitoring" indicates the overload of the A/D
-
Converter.
External Baseband Signal - Baseband Input