![Rohde & Schwarz SMBV100A Operating Manual Download Page 270](http://html1.mh-extra.com/html/rohde-and-schwarz/smbv100a/smbv100a_operating-manual_1477735270.webp)
Instrument Function
R&S
®
SMBV100A
255
Operating Manual 1176.8016.02 ─ 17
FM Deviation
Sets the modulation deviation in Hz.
The deviation is given as the absolute deviation in relation to the set center frequency.
For example, a deviation of 5
MHz at a center frequency of 1
GHz leads to a frequency
modulation between 995 MHz and 1005 MHz.
The maximum deviation depends on the RF frequency and the modulation mode (see
data sheet).
Note that you can set a deviation that is too high for a specific RF frequency, or set an
RF frequency outside of the adjustable range of the deviation. In both cases, the instru-
ment sets the maximum deviation and displays an error message.
In "Int + Ext" modulation source mode, the instrument devides the deviation into half for
each source.
Remote command:
[:SOURce<hw>]:FM:INTernal:DEViation
[:SOURce<hw>]:FM:EXTernal:DEViation
LF Gen Freq
Sets the frequency of the LF generator.
This setting affects all analog modulations which use the LF generator as the internal
modulation source.
Remote command:
LF Gen Shape
Selects the waveform shape of the LF signal.
Note:
The installed hardware determines the available settings. Use the
dialog to check the hardware the instrument is equipped with.
For information on the required hardware revision, refer to the release notes.
Remote command:
FM Sensitivity
Displays the input sensitivity of the externally applied modulation signal at the EXT
MOD input in Hz/V.
The modulation deviation entered with
is achieved with 1 Volt (= U
peak
) of
the input signal.
Note:
The input voltage must not exceed 1.1 V
p
otherwise modulation distortions
occur.
RF Block