Remote Control Commands
R&S
®
SMBV100A
810
Operating Manual 1176.8016.02 ─ 17
Example:
FREQ:STAR
sets the start frequency to 100 MHz.
FREQ:STOP
sets the stop frequency to 500 MHz.
SWE:SPAC LIN
sets linear sweep spacing.
SWE:POIN 401
sets 401 sweep steps for linear sweep spacing. The sweep step
width (
STEP
) is automatically set to 1 MHz.
[:SOURce<hw>]:SWEep[:FREQuency]:RETRace
<State>
Activates that the signal changes to the start frequency value while it is waiting for the
next trigger event.
You can enable this feature, when you are working with sawtooth shapes in sweep
mode "Single" or "External Single".
Parameters:
<State>
0 | 1 | OFF | ON
*RST:
0
Example:
TRIG0:SWE:SOUR SING
FREQ:MODE SWE
SWE:SHAP SAWT
SWE:RETR ON
activates retrace function, i.e. the frequency changes to the
value at start frequency while waiting for the next trigger event.
Manual operation:
See
" Retrace - RF Frequency Sweep "
[:SOURce<hw>]:SWEep[:FREQuency]:RUNNing?
Queries the current state of the frequency sweep mode.
Return values:
<State>
0 | 1 | OFF | ON
Example:
SWE:RUNN?
Response
"1"
: signal generation in level sweep active.
Usage:
Query only
[:SOURce<hw>]:SWEep[:FREQuency]:SHAPe
<Shape>
Sets the cycle mode for a sweep sequence (shape).
SOURce Subsystem