![Rohde & Schwarz SMBV100A Operating Manual Download Page 402](http://html1.mh-extra.com/html/rohde-and-schwarz/smbv100a/smbv100a_operating-manual_1477735402.webp)
Instrument Function
R&S
®
SMBV100A
387
Operating Manual 1176.8016.02 ─ 17
Remote command:
[:SOURce<hw>]:BB:ARBitrary:WAVeform:SELect
Waveform Info
Opens a dialog with detailed information about the currently selected waveform.
The waveform information displayed in this dialog depends on whether the selected
waveform file is a multi segment one or not. All parameters are read-only.
For non-multi segment waveform files, the values of the settings can be queried by
means of the Tag commands (see
Chapter 7.4, "Waveform and List Format"
Remote command:
n.a.
Clock Frequency
Displays or enters the ARB output clock rate.
When the waveform file is loaded, this value is automatically set to the clock rate
defined in the file (Clock tag). The user can subsequently alter the value if necessary.
When intending to work with an external clock source, the frequency to be applied
must be entered here.
Remote command:
[:SOURce<hw>]:BB:ARBitrary:CLOCk