Remote Control Commands
R&S
®
SMBV100A
800
Operating Manual 1176.8016.02 ─ 17
Parameters:
<Sync>
0 | 1 | OFF | ON
ON
Generates a single pulse at the beginning of a pulse sequence,
e.g. to synchronize pulse modulation.
OFF
Returns the validity of the RF signal at the output:
1
(high), while the signal settles.
0
(low), when it is stable (valid).
*RST:
OFF
Example:
PULM:OUTP:SYNC ON
uses the signal for synchronizing the pulse modulation.
Options:
R&S SMBV-K23 (Pulse Generator)
Manual operation:
See
"Use SIGNAL VALID as Pulse Sync"
[:SOURce<hw>]:PULM:PERiod
<Period>
Sets the period of the generated pulse. The period determines the repetition frequency
of the internal signal.
Parameters:
<Period>
float
Range:
5 us | 20 ns to 100 s
Increment: 1us | 5 ns
*RST:
10 us
Example:
PULM:PER 220 us
the pulse period is 220 us.
Options:
R&S
SMBV-K23 (Pulse Generator)
Manual operation:
See
" Pulse Period - Pulse Generator "
[:SOURce<hw>]:PULM:POLarity
<Polarity>
Sets the polarity between modulating and modulated signal. This command is effective
only for an external modulation signal.
Parameters:
<Polarity>
NORMal | INVerted
NORMal
The RF signal is suppressed during the pulse pause.
INVerted
The RF signal is suppressed during the pulse.
*RST:
NORMal
Example:
PULM:SOUR EXT
selects the external modulation source.
SOURce Subsystem