![Rohde & Schwarz SMBV100A Operating Manual Download Page 671](http://html1.mh-extra.com/html/rohde-and-schwarz/smbv100a/smbv100a_operating-manual_1477735671.webp)
Remote Control Commands
R&S
®
SMBV100A
656
Operating Manual 1176.8016.02 ─ 17
●
use the default enabled state, if a common external trigger event is provided for the
master and the slave instruments.
Parameters:
<Output>
0 | 1 | OFF | ON
ON
The signal calculation starts simultaneously with the external
trigger event but because of the instrument’s processing time
the first samples are cut off and no signal is outputted. After
elapsing of the internal processing time, the output signal is syn-
chronous to the trigger event.
OFF
The signal output begins after elapsing of the processing time
and starts with sample 0, i.e. the complete signal is outputted.
This mode is recommended for triggering of short signal sequen-
ces with signal duration comparable with the processing time of
the instrument.
*RST:
1
Example:
BB:DM:TRIG:SOUR EXT
sets external triggering.
BB:DM:TRIG:EXT:SYNC:OUTP ON
enables synchrounous output to external trigger
Manual operation:
See
"Sync. Output to External Trigger"
[:SOURce<hw>]:BB:DM:TRIGger:OUTPut:DELay:FIXed
<Fixed>
The command restricts the marker delay setting range to the dynamic range. In this
range the delay can be set without restarting the marker and signal. If a delay is
entered in setting ON but is outside this range, the maximum possible delay is set and
an error message is generated.
The numeric suffix in
OUTPut
has no significance for this command, since the setting
always affects every marker.
Parameters:
<Fixed>
0 | 1 | OFF | ON
*RST:
0
Example:
BB:DM:TRIG:OUTP:DEL:FIX ON
restricts the marker signal delay setting range to the dynamic
range.
Manual operation:
See