[:SOURce<hw>]:PULM:DOUBle:STATe
<State>
Activates double pulse generation. The two pulses are generated in one pulse period.
Parameters:
<State>
0 | 1 | OFF | ON
*RST:
0
Example:
PULM:DOUB:STAT ON
double-pulse mode is enabled.
Options:
R&S SMBV-K23 (Pulse Generator)
[:SOURce<hw>]:PULM:DOUBle:WIDTh
<Width>
Sets the width of the second pulse in case of double pulse generation.
Parameters:
<Width>
float
Range:
10 ns to 100 s
Increment: 10 ns
*RST:
3 us
Example:
PULM:DOUB:WIDT 33 us
sets a width of 33 us for the second pulse.
Options:
R&S
SMBV-K23 (Pulse Generator)
Manual operation:
See
" Double Pulse Width - Pulse Generator "
[:SOURce<hw>]:PULM:MODE
<Mode>
Sets the mode of the pulse generator.
Parameters:
<Mode>
SINGle | DOUBle
SINGle
Enables single pulse generation.
DOUBle
Enables double pulse generation. The two pulses are generated
in one pulse period.
*RST:
SINGle
Example:
PULM:MODE DOUB
enables double pulse generation.
Options:
R&S SMBV-K23 (Pulse Generator)
Manual operation:
See
" Pulse Mode - Pulse Generator "
[:SOURce<hw>]:PULM:OUTPut:SYNC[:STATe]
<Sync>
Configures the signal at the SIGNAL VALID connector.
SOURce Subsystem