![Rohde & Schwarz SMBV100A Operating Manual Download Page 257](http://html1.mh-extra.com/html/rohde-and-schwarz/smbv100a/smbv100a_operating-manual_1477735257.webp)
Instrument Function
R&S
®
SMBV100A
242
Operating Manual 1176.8016.02 ─ 17
Further functions of the R&S
SMBV related to R&S NRP-Z power sensors are:
●
Acquisition of level correction data, see
Chapter 5.5.9, "User Correction"
on page 232.
The acquired level correction data is used to create and activate lists in which level
correction values predefined by the user are freely assigned to RF frequencies.
Correction is performed by the user-defined table values being added to the output
level for the respective RF frequency.
●
NRP-Z Level Control, see
Chapter 5.5.8, "NRP-Z Level Control"
Note that "NRP-Z Power Viewer" automatically disables "NRP-Z Level Control",
and vice versa.
●
The software version of the connected power sensor can be retrieved by means of
the remote control command
Chapter 5.2.3.5, "NRP-Z Info/Update"
on page 129 dialog to update the
sensor software.
"NRP-Z Power Viewer" automatically disables
NRP-Z Power Viewer Settings
► To access the dialog for configuring the RF signal level, perform one of the follow-
ing:
● Select "RF/A Mod > config... > RF Measurement > NRP-Z Power Viewer".
● Press the MENU key and select "RF/A Mod > RF Measurement > NRP-Z
Power Viewer".
The dialog shows the settings and measurement values of the sensor selected in
the field next to the connector symbol. For indicating the parameters of another
sensor, switch to the respective sensor in the selection list.
RF Block