Remote Control Commands
R&S
®
SMBV100A
696
Operating Manual 1176.8016.02 ─ 17
Example:
BB:MCCW:CARR:LIST:POW -65 dB, -30 dB, -50
dB,...
sets the power of carrier 0 to -65 dB, carrier 1 to -30 dB and so
on.
BB:MCCW:CARR:LIST:POW? 2,3
queries the power of carrier 1 and 2.
Response:
-30,-50
Manual operation:
See
[:SOURce<hw>]:BB:MCCW:CARRier:LIST:STATe
<Stat0[,Stat1..]>]
[:SOURce<hw>]:BB:MCCW:CARRier:LIST:STATe?
<Start>, <Count>
Switches the carrier on or off with the aid of a value list. The first value in the list is
assigned to the carrier with index 0, the second value to the carrier with index 1, etc.
The maximum length corresponds to the maximum number of multi carriers. There is
no need to enter all the values every time. Values not set by the value list are set with
the default values provided they have already been explicitly set by a previous com-
mand. If this is the case, the values continue to apply until overwritten.
Setting parameters:
<Stat0[,Stat1..]>
ON | OFF[,ON | OFF,...]
Query parameters:
<Start>
integer
Range:
0 to lastCarrier
<Count>
integer
Range:
1 to lastCarrier
Return values:
<State[,State..]>
select
Example:
BB:MCCW:CARR:LIST:STAT 1,0,1,1,0,0....
switches carrier 0 on, carrier 1 off, etc.
BB:MCCW:CARR:LIST:STAT? 2,4
queries the states of carrier 2, 3, 4 and 5.
Response:
0,1,1,0
Manual operation:
See
[:SOURce<hw>]:BB:MCCW:CARRier:PHASe
<CarrierIndex>, <Phase>
Sets the start phase of the selected carrier.
The phase settings are only valid if optimization of the crest factor is disabled
(
:SOURce:BB:MCCW:CFACtor:MODE OFF
).
Parameters:
<CarrierIndex>
integer
Range:
0 to lastCarrier
SOURce Subsystem