Instrument Function
R&S
®
SMBV100A
272
Operating Manual 1176.8016.02 ─ 17
"Ext Triggered"
The pulse generator signal is triggered by an external trigger event.
Example: Generation of pulse signals using trigger modes Ext
Triggered (Single Pulse)
The measurement is performed using a 6-dB-attenuator.
Parameter
Value
Trigger Mode
Ext Triggered
Double Pulse State
Off
External Trigger Input Slope
Positive
Pulse Delay
100 ns
Pulse Width
100 ns
Pulse Period
-
Channel 1 = Indicates the external trigger signal. The positive slope is active.
Channel 2 = Indicates the sync signal. The sync signal starts after a trigger delay of
typically 50 ns (see specifications).
Channel 3 = Indicates the pulse signal. The pulse signal starts after the set pulse delay
of 100 ns.
Note:
The sync signal is output at the PULSE SYNC connector, and
the pulse signal at the PULSE VIDEO connector.
Example: Generation of pulse signals using trigger modes Ext
Triggered (Double Pulse)
The measurement were made using a 6-dB-attenuator.
Parameter
Value
Trigger Mode
Ext Triggered
Double Pulse State
On
External Trigger Input Slope
Negative
Pulse Width
100 ns
RF Block