![Rohde & Schwarz SMBV100A Operating Manual Download Page 700](http://html1.mh-extra.com/html/rohde-and-schwarz/smbv100a/smbv100a_operating-manual_1477735700.webp)
Remote Control Commands
R&S
®
SMBV100A
685
Operating Manual 1176.8016.02 ─ 17
Manual operation:
See
[:SOURce<hw>]:BB:DM:SETTing:STORe:FAST
<Fast>
Determines whether the instrument performs an absolute or a differential storing of the
settings.
Enable this function to accelerate the saving process by saving only the settings with
values different to the default ones.
Note:
This function is not affected by the "Preset" function.
Parameters:
<Fast>
0 | 1 | OFF | ON
*RST:
1
Manual operation:
See
[:SOURce<hw>]:BB:DM:STANdard:ULISt:CATalog?
Reads out the files with Digital Standard settings in the default directory. The default
directory is set using command
MMEM:CDIRectory
. Only files with the file extension
*.dm_stu
will be listed.
Return values:
<Catalog>
string
Example:
MMEM:CDIR '/var/user/dig_mod'
sets the default directory.
BB:DM:STAN:ULIS:CAT?
reads out all the files with Digital Standard settings in the default
directory.
Response:
'DM_QAM'
the file
DM_QAM
with Digital Standard settings is available.
Usage:
Query only
Manual operation:
See
[:SOURce<hw>]:BB:DM:STANdard:ULISt:DELete
<Filename>
Deletes the selected file with Digital Standard settings. The directory is set using com-
mand
MMEM:CDIRectory
. A path can also be specified, in which case the files in the
specified directory are read. The file extension may be omitted. Only files with the file
extension
*.dm_stu
will be deleted.
Setting parameters:
<Filename>
string
Example:
BB:DM:STAN:ULIS:DEL 'DM_QAM'
deletes file 'DM_QAM'.
Usage:
Setting only