![Rohde & Schwarz SMBV100A Operating Manual Download Page 774](http://html1.mh-extra.com/html/rohde-and-schwarz/smbv100a/smbv100a_operating-manual_1477735774.webp)
Remote Control Commands
R&S
®
SMBV100A
759
Operating Manual 1176.8016.02 ─ 17
Note:
●
If signal Source "Internal" is set, the instrument performs the analog modulations
(AM/FM/
ɸ
M/PM) with this freqeuncy.
●
In sweep mode (
), the frequency is coupled with the
sweep frequency.
Parameters:
<Frequency>
float
Range:
full frequency range
Increment: see the data sheet: Modulation sources > Resolu-
tion of frequency setting
*RST:
1E3
Example:
LFO:FREQ 5kHz
sets the frequency of the LF signal to 5 kHz
Manual operation:
See
[:SOURce<hw>]:LFOutput:FREQuency:MANual
<Manual>
Determines the frequency and triggers the next sweep step manually in
, and
.
Note:
You can select any frequency within the setting range. The range is defined with
and
Parameters:
<Manual>
float
Range:
full frequency range
Increment: see the data sheet: Modulation sources > Internal
modulation generator > Resolution of frequency
setting
*RST:
1000
Example:
LFO:SWE:MODE MAN
sets the "Step" sweep mode.
LFO:FREQ:MAN 5 kHz
sets an LF frequency of 5 kHz for the next step in the "Step"
sweep mode.
LFO:FREQ:MODE SWE
sets the LF Sweep mode. An LF frequency of 5 kHz is output.
LFO:FREQ:MAN 5.1 kHz
triggers the next sweep step with a frequency of 5.1 kHz.
Manual operation:
See
[:SOURce<hw>]:LFOutput:FREQuency:MODE
<Mode>
Sets the instrument operating mode, and determines the commands to be used for fre-
quency settings.
SOURce Subsystem