![Rohde & Schwarz SMBV100A Operating Manual Download Page 485](http://html1.mh-extra.com/html/rohde-and-schwarz/smbv100a/smbv100a_operating-manual_1477735485.webp)
Remote Control Basics
R&S
®
SMBV100A
470
Operating Manual 1176.8016.02 ─ 17
Use the
>R
to enable the LOCAL key if it is locked.
6.2.3 Examples
This sections provides examples for setting up the remote control connection, and
starting a remote control session.
This section assumes basic knowledge of programming and operation of the controller.
A description of the interface commands can be obtained from the corresponding man-
uals.
6.2.3.1
Remote Control over GPIB
The program example in this section is written in VISUAL BASIC. A condition for pro-
gramming in VISUAL BASIC is that the modules NIGLOBAL (
Niglobal.bas
) and
VBIB32 (
Vbib_32.bas
) are added to the projects.
Drivers for instrument, e.g. IVI-COM and LabVIEW drivers, are available in the down-
load area of the product website (
http://www.rohde-schwarz.com/en/products/
test_and_measurement/product_categories/signal_generation/
).
Starting a remote control session over GPIB
As a prerequisite, the GPIB address of the instrument, which is factory-set to 28, must
not have been changed.
1. Connect instrument and controller using GPIB cable and switch them on.
2. Execute following commands on the controller:
a) Open port to the instrument
CALL IBFIND("DEV1", generator%)
b) Inform controller about instrument address
CALL IBPAD(generator%, 28)
c) Reset instrument
CALL IBWRT(generator%, "*RST;*CLS")
d) Set instrument to new address
CALL IBWRT(generator%, "SYST:COMM:GPIB:ADDR 18")
e) Inform controller about new address
CALL IBPAD(generator%, 18)
The GPIB address of the instrument is changed.
3. To return to manual operation sent
CALL IBLOC (generator%)
or press the
LOCAL key at the front panel.
Starting a Remote Control Session