Remote Control Commands
R&S
®
SMBV100A
695
Operating Manual 1176.8016.02 ─ 17
Setting parameters:
<Phas0[,Phas1..]>
float
Range:
0 to 360
Increment: 0.01
*RST:
0
Default unit: DEG
Query parameters:
<Start>
integer
Range:
0 to lastCarrier
<Count>
integer
Range:
1 to lastCarrier
Return values:
<Phas[,Phas..]>
float
Example:
BB:MCCW:CARR:LIST:PHAS 1,20,30,40,50,60
sets a start phase for carriers 0, 1, 2, 3 and 4.
BB:MCCW:CARR:LIST:PHAS? 2,3
queries the phase of carrier 1, 2 and 3.
Response:
20,30,40
Manual operation:
See
[:SOURce<hw>]:BB:MCCW:CARRier:LIST:POWer
<Pow0[,Pow1..]>]
[:SOURce<hw>]:BB:MCCW:CARRier:LIST:POWer?
<Start>, <Count>
Sets the power of the carrier with the aid of a value list. The first value in the list is
assigned to the carrier with index 0, the second value to the carrier with index 1, etc.
The maximum length corresponds to the maximum number of multi carriers. There is
no need to enter all the values every time. Values not set by the value list are set with
the default values provided they have already been explicitly set by a previous com-
mand. If this is the case, the values continue to apply until overwritten.
Setting parameters:
<Pow0[,Pow1..]>
float
Increment: 0.01
Query parameters:
<Start>
integer
Range:
0 to lastCarrier
<Count>
integer
Range:
1 to lastCarrier
Return values:
<Pow[,Pow..]>
float
SOURce Subsystem