Remote Control Commands
R&S
®
SMBV100A
715
Operating Manual 1176.8016.02 ─ 17
Parameters:
<Mode>
NONE | MASTer | SLAVe
NONE
The instrument is working in stand-alone mode.
MASTer
The instrument provides all connected instrument with its syn-
chronization (including the trigger signal) and reference clock
signal.
SLAVe
The instrument receives the synchronization and reference clock
signal from another instrument working in a master mode.
*RST:
NONE
Example:
BB:MCCW:CLOC:SYNC:MODE MAST
the instrument is configured to work as a master one.
Manual operation:
See
[:SOURce<hw>]:BB:MCCW:CLOCk:SYNChronization:EXECute
Performs automatically adjustment of the instrument's settings required for the syn-
chronization mode, set with the command
.
Example:
BB:MCCW:CLOC:SYNC:MODE MAST
the instrument is configured to work as a master one.
BB:MCCW:CLOC:SYNC:EXEC
all synchronization's settings are adjusted accordingly.
Usage:
Event
Manual operation:
See
"Set Synchronization Settings"
7.16.3.7
SOURce:BB:PROGress Subsystem General Commands
In the R&S
SMBV, some calculation processes may take longer time. While operating
the instrument manually, you can observe the status of an initiated process by the busy
indicator. The following commands fulfill the same task in the remote control operation.
Example: Querying the status of the Create Waveform file process
The following is an example on how to use these commands to retrieve information
about how many percent of the initiated process are completed.
:SOURce1:BB:EUTRa:SETTing:TMOD:DL "E-TM1_1__15MHz"
:SOURce1:BB:EUTRa:SLENgth 100
:SOURce1:BB:PROGress:MCODer?
// 100 (task compleated)
:SOURce1:BB:EUTRa:STATe ON
:BB:PROGress:MCODer?