Remote Control Commands
R&S
®
SMBV100A
649
Operating Manual 1176.8016.02 ─ 17
Example:
BB:DM:SOUR PATT
BB:DM:PATT #B01110111010101010,17
[:SOURce<hw>]:BB:DM:PRBS[:LENGth]
<Length>
The command defines the length of the pseudo-random sequence in accordance with
the following equation:
Length = (2^
Length
) - 1
Parameters:
<Length>
9 | 11 | 15 | 16 | 20 | 21 | 23 | PN9 | PN11 | PN15 | PN16 |
PN20 | PN21 | PN23
*RST:
9
Example:
BB:DM:SOUR PRBS
The internal pseudo-random generator is used as the data
source.
BB:DM:PRBS 9
An internal pseudo-random sequence of 511 bits will be gener-
ated.
Manual operation:
See
[:SOURce<hw>]:BB:DM:PRESet
Sets the default settings for digital modulation (
*RST
values specified for the com-
mands).
Not affected is the state set with the command
SOURce<hw>:BB:DM:STATe
Example:
SOURce1:BB:DM:PRESet
Usage:
Event
Manual operation:
See
[:SOURce<hw>]:BB:DM:SOURce
<Source>
The command selects the data source.
SOURce Subsystem