Remote Control Commands
R&S
®
SMBV100A
596
Operating Manual 1176.8016.02 ─ 17
Example:
SOURce:AWGN:MODE ONLY
SOURce:AWGN:STATe ON
selects the generation of a pure noise and activates it
Manual operation:
See
[:SOURce<hw>]:AWGN:POWer:CARRier
<Carrier>
available for Additive Noise and CW Interferer (
SOUR:AWGN:MODE ADD|CW
) modes
Sets/queries the carrier or signal power depending on the selected reference mode.
●
SOUR:AWGN:POW:RMOD CARR
Sets the carrier power. The power of the noise signal is derived from the entered
C/N value.
●
SOUR:AWGN:POW:RMOD NOIS
queries the carrier power which is derived from the entered C/N value. The noise
power is set with command
SOUR:AWGN:POW:NOISe
.
Parameters:
<Carrier>
float
Increment: 0.01
*RST:
0
Default unit: dBm
Example:
AWGN:DISP:MODE RF
sets the display mode to RF
AWGN:POW:CARR?
queries the carrier power
Manual operation:
See
[:SOURce<hw>]:AWGN:POWer:MODE
<Mode>
In Additive Noise (
SOUR:AWGN:MODE ADD
) mode, selects the mode for setting the
noise power.
Parameters:
<Mode>
CN | SN | EN
CN|SN
The noise power is set on the basis of the value entered for the
carrier/noise or signal/noise ratio
(
SOURce:AWGN:CNRatio|SNRatio
).
Whether a selection of C/N or S/N is enabled, depends on the
selected Display Mode (
AWGN:DISP:MODE BB|RF
)
EN
The noise power is set on the basis of the value entered for the
ratio of bit energy to noise power density (
AWGN:ENR
).
*RST:
CN|SN
SOURce Subsystem