![Rohde & Schwarz SMBV100A Operating Manual Download Page 345](http://html1.mh-extra.com/html/rohde-and-schwarz/smbv100a/smbv100a_operating-manual_1477735345.webp)
Instrument Function
R&S
®
SMBV100A
330
Operating Manual 1176.8016.02 ─ 17
The instruments have to be configured and connected as described in the following
sections.
This straightforward method ensure the simultaneous and synchronous signal genera-
tion in both the baseband and the RF part. However, to ensure even better phase
coherence of the signal at the RF outputs, the LO Coupling feature of the R&S
SMBV
should be used. For more information about this feature, see
In the R&S
SMBV, the parameters required to achieve a “perfectly synchronized” state
have to be gathered only once; they are stable in the time are not influenced by turning
the instrument on and off. Refer to Application Note 1GP67 for detailed information on
how to adjust the required settings.
Connecting Several R&S SMBVs for Precise Synchronous Signal Generation
For working in master-slave synchronous mode, connect the instruments as follow:
1. Connect the instruments as a daisy chain, i.e. connect the inputs of each further
instrument to the outputs of the previous one.
2. Connect the output CLK OUT of the master instrument to the signal input CLK IN
of the slave one.
Note:
From firmware version 2.05.269 on, the trigger information is modulated on
the clock signal and no additional cabling is necessary. For older firmware versions
however, a connection between the output MARKER 1 of the master instrument
and the signal input TRIG of the slave one is required.
3. To ensure phase coherence of the RF signals, connect the output of the local oscil-
lator LO OUT of the master instrument to the signal input LO IN of the first slave
one.
The LO Coupling feature is available only for instruments equipped with the option
R&S
SMBV–B90.
4. Provide a reference signal for the instruments, i.e. connect the output REF OUT of
the master instrument to the signal input REF IN of the first slave one or use a
common external reference signal.
This connection is not necessary, if the local oscillators are connected.
5. Connect the power combiner and the spectrum analyzer. The connections are nec-
essary only during the adjustment phase.
6. Avoid unnecessary cable lengths and branching points.
The figure below shows the cabling of two or more R&S SMBVs for working in master-
slave synchronous mode.
Baseband Signal - Baseband Block