![Rohde & Schwarz SMBV100A Operating Manual Download Page 754](http://html1.mh-extra.com/html/rohde-and-schwarz/smbv100a/smbv100a_operating-manual_1477735754.webp)
Remote Control Commands
R&S
®
SMBV100A
739
Operating Manual 1176.8016.02 ─ 17
Example:
FREQ:MODE SWE
sets the SWEep mode.
Manual operation:
See
[:SOURce<hw>]:FREQuency:MULTiplier
<Multiplier>
Sets the value for the multiplication factor of a subsequent downstream instrument.
Parameters:
<Multiplier>
float
Range:
1 to dynamic
Increment: 0.001
*RST:
1
Example:
FREQ:MULT 1
sets the multiplication factor to 1.
Manual operation:
See
[:SOURce<hw>]:FREQuency:OFFSet
<Offset>
Sets the frequency offset of a downstream instrument, for example a mixer.
If you have specified an
OFFSet
and / or a
MULTiplier
factor, the actual frequency
at the RF output does not change, but rather the value queried with
according to the following formula:
f
FREQ
= f
RFout
* f
MULTiplier
+ f
OFFSet
Parameters:
<Offset>
float
Increment: 0.01
*RST:
0
Example:
FREQ:OFFS 500kHz
sets the frequency offset to 500 kHz.
Manual operation:
See
[:SOURce<hw>]:FREQuency:SPAN
<Span>
Determines the extent of the frequency sweep range. This setting in combination with
the center frequency setting (
[:SOURce<hw>]:FREQuency:CENTer
) defines the
sweep range.
This parameter is related to the start and stop frequencies. If you change the fre-
quency, the span changes accordingly.
f
SPAN
= f
STOP
- f
STARt
f
STARt
> f
STOP
is permitted.
SOURce Subsystem