Remote Control Commands
R&S
®
SMBV100A
653
Operating Manual 1176.8016.02 ─ 17
Example:
BB:DM:CLOC:SOUR EXT
selects the external clock source. The clock is supplied via the
CLOCK connector.
BB:DM:CLOC:MODE MSYM
selects clock type Multiple Symbols, i.e. the supplied clock has a
rate which is a multiple of the symbol rate.
BB:DM:CLOC:MULT 12
the multiplier for the external clock rate is 12.
Manual operation:
See
[:SOURce<hw>]:BB:DM:CLOCk:SOURce
<Source>
The command selects the source for the digital modulation clock.
Parameters:
<Source>
INTernal | EXTernal
INTernal
The internal clock reference is used.
EXTernal
The external clock reference is supplied to the CLOCK connec-
tor.
*RST:
INTernal
Example:
BB:DM:CLOC:SOUR INT
selects the internal clock reference.
Manual operation:
See
[:SOURce<hw>]:BB:DM:CLOCk:SYNChronization:MODE
<Mode>
Selects the synchronization mode
This parameter is used to enable generation of very precise synchronous signal of sev-
eral connected R&S SMBVs.
Note
: If several instruments are connected, the connecting cables from the master
instrument to the slave one and between each two consecutive slave instruments must
have the same length and type. Avoid unnecessary cable length and branching points.
SOURce Subsystem