![Rohde & Schwarz SMBV100A Operating Manual Download Page 841](http://html1.mh-extra.com/html/rohde-and-schwarz/smbv100a/smbv100a_operating-manual_1477735841.webp)
Remote Control Commands
R&S
®
SMBV100A
826
Operating Manual 1176.8016.02 ─ 17
:SYSTem:DLOCk
<DispLockStat>
Disables the display, or enables it again (
OFF
).
The command disables also the front panel keyboard of the instrument including the
LOCAL key.
Parameters:
<DispLockStat>
0 | 1 | OFF | ON
*RST:
n.a. (factory preset: 0)
Example:
SYST:DLOC ON
Locks the display. To unlock the display
SYST:DLOC OFF
.
Manual operation:
See
:SYSTem:KLOCk
<State>
Keyboard lock disables the front panel keyboard of the instrument including the LOCAL
key, or enables it again (
OFF
).
The command disables also the front panel keyboard of the instrument including the
LOCAL key.
Parameters:
<State>
0 | 1 | OFF | ON
*RST:
n.a. (factory preset: 0)
Example:
SYST:KLOC ON
Locks the front panel and external controls. To enable the con-
trols, set
SYST:KLOC OFF
.
Manual operation:
See
:SYSTem:ULOCk
<Mode>
Locks or unlocks the user interface of the instrument.
Parameters:
<Mode>
ENABled | DONLy | DISabled | VNConly
ENABled
Unlocks the display and all controls for the manual operation.
DONLy
Locks the controls for the manual operation of the instrument.
The display shows the current settings.
DISabled
Locks the controls for the manual operation, and enables remote
operation over VNC. The display shows the current settings.
VNConly
Locks the display and all controls for the manual operation.
*RST:
n.a. (factory preset: ENABled)
SYSTem Subsystem