R&S
®
ZVA/ZVB/ZVT
Reflection Measurements
1145.1084.12 2.5
E-1
The next dialog of the calibration wizard shows that only a single calibration standard
needs to be measured.
5. Click
the
box
Short (m)...
to initiate the measurement of the connected short
standard.
The analyzer performs a calibration sweep and displays a message box with a progress
bar. After completing the sweep the analyzer generates a short sound and a green
checkmark appears in the checkbox.
6. Click
Apply
to close the wizard, calculate and store the system error correction
data and apply them to the current measurement.
7. Remove the short standard and connect the DUT again.