Channel Menu
R&S
®
ZVA/ZVB/ZVT
1145.1084.12 4.142
E-6
•
Name
adds a column to assign a name to each segment. A segment name is a string that may
contain letters, numbers and special characters.
•
Power
defines the internal source
Power
for each individual sweep segment.
•
Meas Bandwidth
defines the
Meas Bandwidth
for each individual sweep segment.
•
Selectivity
defines the
Selectivity
of the IF filter used for each sweep segment.
•
Spur Avoid
defines whether the analyzer measures the segment with a local oscillator
frequency LO below or above the RF input frequency. The parameter replaces the
Spurious
Avoidance
settings for a particular segment.
•
Time
defines the sweep time for each segment. The default configuration for a new segment is equal
to the Sweep Time settings.
When
Time
is checked two new columns appear in the table. The first column reads
Swp Tm
or
Pnt Dly,
depending on the selected radio button below the
Time
checkbox. The second column
reads
A(uto)
and is used to activate automatic sweep time setting.
•
A(uto)
minimizes the sweep time. If this option is checked, the columns
Swp Tm
or
Meas. Dly
can not be edited.
Swp Tm
indicates the estimated sweep time, taking into
account the current measurement settings, the
Meas. Dly
is 0 s. The segment sweep
time and point delay values are maintained until changed explicitly if
Auto
is switched
off.
•
Swp Tm or Segment Sweep Time
is the total measurement time for the sweep
segment. The minimum segment sweep time to be set is equal to the estimated value
in the
Auto
mode.
•
Meas. Delay
sets a delay time allowing the DUT to settle before the hardware settings
of the analyzer are changed and a new partial measurement is started.
Changing the
Meas. Dalay
modifies the
Swp Tm
and vice versa.
Remote control:
[SENSe<Ch/Tr>:]SEGMent<Seg>:POWer[:LEVel]
[SENSe<Ch/Tr>:]SEGMent<Seg>:BWIDth[RESolution]
[SENSe<Ch/Tr>:]SEGMent<Seg>:BWIDth[RESolution]:SELect
[SENSe<Ch/Tr>:]SEGMent<Seg>:DEFine
[SENSe<Ch/Tr>:]SEGMent<Seg>:INSert
[SENSe<Ch/Tr>:]SEGMent<Seg>:SWEep:TIME
[SENSe<Ch/Tr>:]SEGMent<Seg>:SWEep:DWELl
Point List, Segment Import and Export
The buttons on the right side below the table in the
Define Segments
dialog are used to retrieve the
position of the defined sweep points import and export limit line data.
•
Show Point List
opens a list of all active sweep points and their channel settings.
Содержание 1145.1010.04/05/06
Страница 10: ......
Страница 20: ......
Страница 22: ......
Страница 24: ...Contents of Chapter 1 R S ZVA 1145 1090 62 I 1 2 E 1 Windows XP Embedded 1 22 Firmware Update 1 22...
Страница 48: ......
Страница 70: ......
Страница 72: ......
Страница 90: ......
Страница 92: ......
Страница 108: ......
Страница 156: ......
Страница 162: ......
Страница 406: ...Display Menu R S ZVA ZVB ZVT 1145 1084 12 4 244 E 6 Stack Tile Horizontally Tile Vertically...
Страница 450: ...Status Reporting System R S ZVA ZVB ZVT 1145 1084 12 5 18 E 1...
Страница 462: ......
Страница 766: ......
Страница 772: ......
Страница 792: ......
Страница 794: ......
Страница 808: ......