[SENSe<Ch>:]AVERage
[SENSe<Ch>:]AVERage...
This subsystem sets sweep averaging parameters. The sweep average is a noise-reduction technique
which consists of calculating each measurement point as an average of the same measurement point
over several consecutive sweeps.
[SENSe<Ch>:]AVERage:CLEar
Starts a new average cycle, clearing all previous results and thus eliminating their effect on the new
cycle.
<Ch>
Channel number. If unspecified the numeric suffix is set to 1.
*RST value
–
SCPI, Command
Types
Confirmed, no query
Example:
AVER:COUN 15; AVER ON
Set the average factor for channel 1 to 15 and enable the
sweep average.
AVER:COUN 5; CLE
Reduce the average factor and restart the average.
[SENSe<Ch>:]AVERage:COUNt <numeric_value>
Defines the number of consecutive sweeps to be combined for the sweep average
(Average Factor)
.
<Ch>
Channel number.
<numeric_value>
Average factor
Range [def. unit]
1 to 1000 [1]
*RST value
10
SCPI, Command Types
Confirmed, command or query
Example:
AVER:COUN 15
Set the average factor for channel 1 to 15.
AVER ON
Enable the sweep average.
Содержание 1145.1010.04/05/06
Страница 10: ......
Страница 20: ......
Страница 22: ......
Страница 24: ...Contents of Chapter 1 R S ZVA 1145 1090 62 I 1 2 E 1 Windows XP Embedded 1 22 Firmware Update 1 22...
Страница 48: ......
Страница 70: ......
Страница 72: ......
Страница 90: ......
Страница 92: ......
Страница 108: ......
Страница 156: ......
Страница 162: ......
Страница 406: ...Display Menu R S ZVA ZVB ZVT 1145 1084 12 4 244 E 6 Stack Tile Horizontally Tile Vertically...
Страница 450: ...Status Reporting System R S ZVA ZVB ZVT 1145 1084 12 5 18 E 1...
Страница 462: ......
Страница 766: ......
Страница 772: ......
Страница 792: ......
Страница 794: ......
Страница 808: ......