Trace Menu
R&S
®
ZVA/ZVB/ZVT
1145.1084.12 4.116
E-6
TTL Out Pass 1/2
Assigns the active trace to the low-voltage (3.3 V) TTL output signals at the
USER CONTROL
connector. To select
TTL Out Pass 1 / 2
, the limit check of the active trace must be switched on.
•
If
TTL Out Pass 1
is selected and the active trace is within limits, then the TTL signal is applied
to pin 13 of the
USER CONTROL
connector.
•
If
TTL Out Pass 2
is selected and the active trace is within limits, then the TTL signal is applied
to pin 14 of the
USER CONTROL
connector.
If the active trace exceeds the limits, then no TTL signal is generated. It is possible to activate both
pass/fail signals for the same trace or assign several traces to a signal.
Extension: Monitoring several traces
If a channel contains several traces, is possible to assign them one after another to each pass/fail
signal. The procedure divides the traces of the channel into four groups that are either assigned to
signal A, to signal B, to both signals, or to none of them.
If several traces with independent limit check are assigned to a pass/fail signal, then the TTL signal is
generated only if all traces are within limits. It is switched off as soon as one trace exceeds the limits.
Application: Graduated limit check
The two pass/fail signals can be used to distinguish three quality levels of a DUT. The test is performed
with a looser and a tighter set of limit lines that are assigned to two traces with identical channel and
trace settings. The limit check for the two traces is monitored by means of the signals
TTL Out Pass 1
/TTL Out Pass 2,
respectively.
•
If the DUT is passed in both limit checks, the quality is good.
•
If the DUT is failed in the limit check with tighter limits, the quality is still sufficient.
•
If the DUT is failed in both limit checks, the quality is poor.
Instead of using two traces, it is possible to consider two groups of traces that are assigned to
TTL Out
Pass 1
and
TTL Out Pass 2
.
Remote control:
CALCulate<Ch/Tr>:LIMit:TTLout<nr>[:STATe] ON | OFF
Horizontal Line
Shows or hides the horizontal line associated to the active trace in a Cartesian diagram area. A
checkmark appears next to the menu item when the horizontal line is shown.
The horizontal line (or display line) is a red line which can be moved to particular trace points in order to
retrieve the response values.
Содержание 1145.1010.04/05/06
Страница 10: ......
Страница 20: ......
Страница 22: ......
Страница 24: ...Contents of Chapter 1 R S ZVA 1145 1090 62 I 1 2 E 1 Windows XP Embedded 1 22 Firmware Update 1 22...
Страница 48: ......
Страница 70: ......
Страница 72: ......
Страница 90: ......
Страница 92: ......
Страница 108: ......
Страница 156: ......
Страница 162: ......
Страница 406: ...Display Menu R S ZVA ZVB ZVT 1145 1084 12 4 244 E 6 Stack Tile Horizontally Tile Vertically...
Страница 450: ...Status Reporting System R S ZVA ZVB ZVT 1145 1084 12 5 18 E 1...
Страница 462: ......
Страница 766: ......
Страница 772: ......
Страница 792: ......
Страница 794: ......
Страница 808: ......