Channel Menu
R&S
®
ZVA/ZVB/ZVT
1145.1084.12 4.202
E-6
A successful calibration will supersede the previous calibration, discarding all previous system
error correction data. To keep older correction data you can transfer them into a
Cal Pool
using the
Calibration Manager.
The system error correction data determined in a calibration procedure are stored on the analyzer. You
can read these correction data using the remote control command
[SENSe<Ch>:]CORRection:CDATa.
You can also replace the correction data of the analyzer by
your own correction data sets.
Select Connectors
The first dialog of the calibration wizard displays a table to select the connectors and calibration kits for
all calibrated physical ports.
The table contains the following rows:
•
Physical Port Number #
The ports (and therefore the number of table rows) are determined by the active calibration type
selected in the
Start Cal
submenu. For a user-defined calibration
(Start Cal
–
Other...)
all
measurement ports of the analyzer are displayed regardless of their port configuration.
Содержание 1145.1010.04/05/06
Страница 10: ......
Страница 20: ......
Страница 22: ......
Страница 24: ...Contents of Chapter 1 R S ZVA 1145 1090 62 I 1 2 E 1 Windows XP Embedded 1 22 Firmware Update 1 22...
Страница 48: ......
Страница 70: ......
Страница 72: ......
Страница 90: ......
Страница 92: ......
Страница 108: ......
Страница 156: ......
Страница 162: ......
Страница 406: ...Display Menu R S ZVA ZVB ZVT 1145 1084 12 4 244 E 6 Stack Tile Horizontally Tile Vertically...
Страница 450: ...Status Reporting System R S ZVA ZVB ZVT 1145 1084 12 5 18 E 1...
Страница 462: ......
Страница 766: ......
Страница 772: ......
Страница 792: ......
Страница 794: ......
Страница 808: ......