SOURce<Ch>:POWer:CORRection...
R&S
®
ZVA/ZVB/ZVT
1145.1084.12 6.272
E-6
<numeric_value>
Stop power of the sweep.
Range [def. unit]
–39.99 dBm to +10 dBm. The exact range depends on the
analyzer model; refer to the data sheet [dBm]. UP and DOWN
increment/decrement the source power in 1-dB steps.
*RST value
0 dBm
SCPI, Command
Types
Confirmed, command or query
Example:
SWE:TYPE POW
Activate a power sweep.
SOUR:POW:STAR -6; STOP 10
Select a power sweep range between –6 dBm and +10 dBm.
Note:
If the stop power entered is greater than the current start power
(
SOURce<Ch>:POWer<Pt>:STARt
), the start power is set to the stop power minus the
minimum power span of 0.01 dB.
SOURce<Ch>:POWer:CORRection
SOURce<Ch>:POWer:CORRection...
This subsystem controls the source power calibration.
Note:
The SOURce<Ch>:POWer... subsystem comprises port-specific and general settings. Port-
specific settings are valid for the port specified by the numeric suffix
<Pt>
(...
:POWER<Pt>:
...).
General settings are valid for all test ports of the analyzer; the port suffix is ignored. Refer to the
description of the individual commands for more information.
SOURce<Ch>:POWer<Pt>:CORRection[:ACQuire] PORT | GENerator | A1 | A2 |
A3 | A4 | ESRC1 | ESRC2[,<port_number>]
Selects the source for the source power calibration, starts and applies the source power calibration.
The command cannot be used unless a power meter is connected via GPIB bus, USB or LAN
interface and configured in the External Power Meters dialog.
<Ch>
Calibrated channel number
<pt>
Calibrated port number. This suffix is ignored because the port
number is specified in the parameter list.
PORT
GENerator
Analyzer port is the source of the calibrated wave.
External generator is the source of the calibrated wave.
The port and generator numbers are specified by means of the
<port_number> parameter.
A1 | A2 | A3 | A4
Direct analyzer port selection. The parameters A1 and A2 are
ZVR-compatible. No additional parameters need to be specified.
ESRC1 | ESRC2
Direct external generator selection. The parameters are ZVR-
compatible. No additional parameters need to be specified.
Содержание 1145.1010.04/05/06
Страница 10: ......
Страница 20: ......
Страница 22: ......
Страница 24: ...Contents of Chapter 1 R S ZVA 1145 1090 62 I 1 2 E 1 Windows XP Embedded 1 22 Firmware Update 1 22...
Страница 48: ......
Страница 70: ......
Страница 72: ......
Страница 90: ......
Страница 92: ......
Страница 108: ......
Страница 156: ......
Страница 162: ......
Страница 406: ...Display Menu R S ZVA ZVB ZVT 1145 1084 12 4 244 E 6 Stack Tile Horizontally Tile Vertically...
Страница 450: ...Status Reporting System R S ZVA ZVB ZVT 1145 1084 12 5 18 E 1...
Страница 462: ......
Страница 766: ......
Страница 772: ......
Страница 792: ......
Страница 794: ......
Страница 808: ......