R&S
®
ZVA/ZVB/ZVT
CALCulate
1145.1084.12 6.113
E-6
CALCulate<Ch/Tr>:TRANsform:TIME:LPASs:DCSParam:CONTinuous
<Boolean>
Determines whether continuous extrapolation for the DC value is enabled.
<Ch/Tr>
Channel number used to identify the active trace
<Boolean>
ON – Continuous extrapolation enabled
OFF – Continuous extrapolation disabled
*RST value
OFF [–]
SCPI, Command
Types
Device-specific, command or query
Example:
See
CALCulate<Ch/Tr>:TRANsform:TIME:LPASs:DCSParam
CALCulate<Ch/Tr>:TRANsform:TIME:LPASs:DCSParam:EXTRapolate
Extrapolates the measured trace towards f = 0 and overwrites the current DC value
(
CALCulate<Ch/Tr>:TRANsform:TIME:LPASs:DCSParam
). The command is relevant for low pass
time domain transforms.
<Ch/Tr>
Channel number used to identify the active trace
Parameters
None
*RST value
–
SCPI, Command
Types
Device-specific, no query (the command causes an event)
Example:
See
CALCulate<Ch/Tr>:TRANsform:TIME:LPASs:DCSParam
CALCulate<Ch/Tr>:TRANsform:TIME:LPFRequency
Calculates the harmonic grid for low pass time domain transforms, keeping the stop frequency and the
number of points.
Use
CALCulate<Ch/Tr>:TRANsform:TIME:LPASs
if you wish to use one of the other
algorithms for calculating the grid.
<Ch/Tr>
Channel number used to identify the active trace
Parameters
None
*RST value
–
SCPI, Command Types
Device-specific, no query (the command causes an event)
Example:
See CALCulate<Ch/Tr>:TRANsform:TIME[:TYPE]
Содержание 1145.1010.04/05/06
Страница 10: ......
Страница 20: ......
Страница 22: ......
Страница 24: ...Contents of Chapter 1 R S ZVA 1145 1090 62 I 1 2 E 1 Windows XP Embedded 1 22 Firmware Update 1 22...
Страница 48: ......
Страница 70: ......
Страница 72: ......
Страница 90: ......
Страница 92: ......
Страница 108: ......
Страница 156: ......
Страница 162: ......
Страница 406: ...Display Menu R S ZVA ZVB ZVT 1145 1084 12 4 244 E 6 Stack Tile Horizontally Tile Vertically...
Страница 450: ...Status Reporting System R S ZVA ZVB ZVT 1145 1084 12 5 18 E 1...
Страница 462: ......
Страница 766: ......
Страница 772: ......
Страница 792: ......
Страница 794: ......
Страница 808: ......