INSTrument...
R&S
®
ZVA/ZVB/ZVT
1145.1084.12 6.164
E-6
Example:
INP2:ATT 10
Set the step attenuator for the wave received at port 2 and for all
channels to 10 dB. The waves at the other test ports are not
affected.
INSTrument...
This subsystem identifies and selects particular resources (SCPI: logical instruments) of the analyzer.
INSTrument:NSELect <Ch>
Selects a channel <Ch> as the active channel.
<Ch>
Number of the channel to be activated. The channel must be created
before using
CONFigure:CHANnel<Ch>:STATe ON
.
Range [def.
unit]
1, 2, ... [–]
*RST value
1
SCPI,
Command
Types
Confirmed, command or query. If all channels have been deleted
CONFigure:CHANnel<Ch>:STATe OFF
the query returns 0.
Example:
CONF:CHAN2:STAT ON; INST:NSEL?
Create channel no. 2 and select it as the active
channel. The query returns 2.
INSTrument:PORT:COUNt?
Returns the number of test ports of the analyzer.
Response
Number of ports (integer number).
Range [def. unit]
1, 2, 3 or 4, depending on the analyzer model.
*RST value
–
SCPI, Command Types
Device-specific, query only.
Example:
INP:PORT:COUN?
Содержание 1145.1010.04/05/06
Страница 10: ......
Страница 20: ......
Страница 22: ......
Страница 24: ...Contents of Chapter 1 R S ZVA 1145 1090 62 I 1 2 E 1 Windows XP Embedded 1 22 Firmware Update 1 22...
Страница 48: ......
Страница 70: ......
Страница 72: ......
Страница 90: ......
Страница 92: ......
Страница 108: ......
Страница 156: ......
Страница 162: ......
Страница 406: ...Display Menu R S ZVA ZVB ZVT 1145 1084 12 4 244 E 6 Stack Tile Horizontally Tile Vertically...
Страница 450: ...Status Reporting System R S ZVA ZVB ZVT 1145 1084 12 5 18 E 1...
Страница 462: ......
Страница 766: ......
Страница 772: ......
Страница 792: ......
Страница 794: ......
Страница 808: ......