SEGM:ADD
Create a new sweep segment no. 1 in channel no. 1 using default
settings and thus –10 dBm internal source power.
SEGM:POW -20
Decrease the power to –20 dBm.
SEGM:POW:CONT OFF
Couple the powers in all segments and reset the power in segment no. 1
to the initial value.
[SENSe<Ch>:]SEGMent<Seg>[:STATe] <Boolean>
Activates or deactivates the sweep segment <Seg>. Sweep points belonging to inactive segments only
are not measured
<Ch>
Channel number. If unspecified the numeric suffix is set to 1.
<Seg>
Sweep segment number.
<Boolean>
ON | OFF - Activates or deactivates the measurement in sweep
segment <Seg>.
*RST value
ON
SCPI, Command
Types
Device-specific, command or query
Example:
SEGM:ADD
Create a new sweep segment no. 1 in channel no. 1 using
default settings.
SEGM OFF
Disable the measurement in the created sweep segment.
[SENSe<Ch>:]SEGMent<Seg>:SWEep:POINts <numeric_value>
Defines the total number of measurement
Points
in sweep segment no. <Seg>.
<Ch>
Channel number 1.
<Seg>
Sweep segment number.
<numeric_value>
Number of points in the segment.
Содержание 1145.1010.04/05/06
Страница 10: ......
Страница 20: ......
Страница 22: ......
Страница 24: ...Contents of Chapter 1 R S ZVA 1145 1090 62 I 1 2 E 1 Windows XP Embedded 1 22 Firmware Update 1 22...
Страница 48: ......
Страница 70: ......
Страница 72: ......
Страница 90: ......
Страница 92: ......
Страница 108: ......
Страница 156: ......
Страница 162: ......
Страница 406: ...Display Menu R S ZVA ZVB ZVT 1145 1084 12 4 244 E 6 Stack Tile Horizontally Tile Vertically...
Страница 450: ...Status Reporting System R S ZVA ZVB ZVT 1145 1084 12 5 18 E 1...
Страница 462: ......
Страница 766: ......
Страница 772: ......
Страница 792: ......
Страница 794: ......
Страница 808: ......