R&S
®
ZVA/ZVB/ZVT
SENSe
1145.1084.12 6.201
E-6
[SENSe<Ch>:]CORRection:COLLect:DELete ['<cal_name>']
Deletes system error correction data generated and stored previously
.
<Ch>
Channel number of the calibrated channel.
'<cal_name>'
Name of the calibration (string parameter) defined together with
calibration type
[SENSe<Ch>:]CORRection:COLLect:METHod:DEFine
. If nothing
is specified the analyzer deletes the last system error correction stored
by means of
[SENSe<Ch>:]CORRection:COLLect:SAVE
.
*RST value
–
SCPI,
Command
Types
Device-specific, no query
Example:
CORR:COLL:METH:DEF 'Test',RSHort,1
Select a one-port normalization at port 1 with a short standard as
calibration type.
CORR:COLL:SEL SHOR,1
Measure a short standard connected to port 1 and store the
measurement results of this standard.
CORR:COLL:SAVE; DEL
Calculate the system error correction data and apply them to the active
channel, then delete the data.
[SENSe<Ch>:]CORRection:COLLect:METHod FRTRans | TOM | TRM | TRL | TNA
| TOSM | FOPort1 | FOPort2 | FOPort12 | FOPTport | ROPTport | REFL1 | REFL2 |
REFL12 | TPORt | ETOM | ETSM
Selects a one-port or two-port calibration type at ports 1/2.
Use the generalized command
[SENSe<Ch>:]CORRection:COLLect:METHod:DEFine
to
select the calibration type for arbitrary analyzer ports or a multiport calibration type.
<Ch>
Channel number of the calibrated channel.
Parameters
Calibration types:
TOM, TRM, TRL, TNA, TOSM, Full One Port, One
Path Two Port, Normalization
(REFL1, REFL2 and REFL12 for one-
port, TPORT for two-port)
, TOM Enhanced, TSM Enhanced
.
The numbers in the parameters denote the analyzer ports. Parameters
for two-port calibration types contain no numbers because the
command is only valid for ports 1 and 2.
*RST value
–
SCPI,
Command
Types
Confirmed (with device-specific calibration types), command or query. If
several calibration types are assigned to channel <Ch>, the query
returns a list of all calibration types.
Example:
See
[SENSe<Ch>:]
CORRection:COLLect:SAVE
.
Содержание 1145.1010.04/05/06
Страница 10: ......
Страница 20: ......
Страница 22: ......
Страница 24: ...Contents of Chapter 1 R S ZVA 1145 1090 62 I 1 2 E 1 Windows XP Embedded 1 22 Firmware Update 1 22...
Страница 48: ......
Страница 70: ......
Страница 72: ......
Страница 90: ......
Страница 92: ......
Страница 108: ......
Страница 156: ......
Страница 162: ......
Страница 406: ...Display Menu R S ZVA ZVB ZVT 1145 1084 12 4 244 E 6 Stack Tile Horizontally Tile Vertically...
Страница 450: ...Status Reporting System R S ZVA ZVB ZVT 1145 1084 12 5 18 E 1...
Страница 462: ......
Страница 766: ......
Страница 772: ......
Страница 792: ......
Страница 794: ......
Страница 808: ......