Remote Control: Introduction
R&S
®
ZVA/ZVB/ZVT
1145.1084.12 5.10
E-1
The key words feature a long form and a short form. Either the short form or the long form can be
entered; other abbreviations are not permitted.
Example:
TRIGger:SOURce EXTernal
TRIG:SOUR EXT
The short form is marked by upper case letters, the long form corresponds to the
complete word. Upper case and lower case notation only serves to distinguish the two
forms in the manual, the instrument itself is case-insensitive.
•
Parameters
Parameters must be separated from the header by a "white space". If several parameters are specified
in a command, they are separated by a comma ",". For a description of the types of parameter, refer to
section Parameters.
Example:
SOURce:GROup 1,1
This command defines a group of measured ports.
•
Numeric suffix
If a device features several functions or features of the same kind, e.g. several channels or test ports,
the desired function can be selected by a suffix added to the command. Entries without suffix are
interpreted like entries with the suffix 1.
Example:
SOURce:GROup2 1,1
This command defines a second group (group no 2) of measured ports.
Structure of a Command Line
A command line may consist of one or several commands. It is terminated by a <New Line>, a <New
Line> with EOI or an EOI together with the last data byte. Visual BASIC automatically produces an EOI
together with the last data byte.
Several commands in a command line must be separated by a semicolon ";". If the next command
belongs to a different command system, the semicolon is followed by a colon.
Example:
CALL IBWRT(device%, "TRIGger:SOURce EXTernal;:SENSe:FREQuency:STARt 1GHZ")
This command line contains two commands. The first command belongs to the
TRIGger
system and
defines the trigger source (external trigger). The second command belongs to the
SENSe
system and
defines the start frequency of the sweep.
If the successive commands belong to the same system, having one or several levels in common, the
command line can be abbreviated. To this end, the second command after the semicolon starts with the
level that lies below the common levels. The colon following the semicolon must be omitted in this case.
Example:
CALL IBWRT(device%, "TRIG:SOUR EXT;:TRIG:TIM 0.1")
This command line is represented in its full length and contains two commands separated from each
other by the semicolon. Both commands are part of the
TRIGger
command system, i.e. they have one
level in common.
When abbreviating the command line, the second command begins with the level below
TRIG
. The
colon after the semicolon is omitted. The abbreviated form of the command line reads as follows:
Содержание 1145.1010.04/05/06
Страница 10: ......
Страница 20: ......
Страница 22: ......
Страница 24: ...Contents of Chapter 1 R S ZVA 1145 1090 62 I 1 2 E 1 Windows XP Embedded 1 22 Firmware Update 1 22...
Страница 48: ......
Страница 70: ......
Страница 72: ......
Страница 90: ......
Страница 92: ......
Страница 108: ......
Страница 156: ......
Страница 162: ......
Страница 406: ...Display Menu R S ZVA ZVB ZVT 1145 1084 12 4 244 E 6 Stack Tile Horizontally Tile Vertically...
Страница 450: ...Status Reporting System R S ZVA ZVB ZVT 1145 1084 12 5 18 E 1...
Страница 462: ......
Страница 766: ......
Страница 772: ......
Страница 792: ......
Страница 794: ......
Страница 808: ......