R&S
®
ZVA/ZVB/ZVT
Channel
Menu
1145.1084.12 4.235
E-6
Auto Length
Adds an electrical length offset to the active test port with the condition that the residual delay of the
active trace (defined as the negative derivative of the phase response) is minimized across the entire
sweep range. If
Delay
is the selected trace format, the entire trace is shifted in vertical direction and
centered around zero. The
Auto Length
corrected trace shows the deviation from linear phase. The
effect of a dispersive connector type (i.e. a waveguide; see
Offset Model
dialog) assigned to the active
test port is taken into account.
The active test port number is defined as the port index in the numerator of the measured quantity. If
the active trace shows an S-parameter S
ij
, then
Auto Length
adds a length offset at port i, which is the
receive port of the analyzer. The added length offset is valid for all traces associated with the active test
port.
Preconditions for Auto Length, effect on measured quantities and exceptions
Auto Length
is enabled if the measured quantity contains the necessary phase information as a
function of the frequency, and if the interpretation of the results is unambiguous:
•
A frequency sweep must be active.
•
The measured quantity must be an S-parameter, ratio, wave quantity, a converted
impedance or a converted admittance.
The effect of
Auto Length
on S-parameters, wave quantities and ratios is to eliminate a linear
phase response as described above. Converted admittances or impedances are calculated
from the corresponding
Auto Length
corrected S-parameters. Y-parameters, Z-parameters and
stability factors are not derived from a single S-parameter, therefore
Auto Length
is disabled.
Auto Length for logical ports
The
Auto Length
function can be used for balanced port configurations as well. If the active test
port is a logical port, then the same length offset is assigned to both physical ports that are
combined to form the logical port. If different length offsets have been assigned to the physical
ports before, they are both corrected by the same amount.
Use
Zero Delay at Marker
to set
the delay at a special trace point to zero.
Remote control:
[SENSe<Ch>:]CORRection:EDELay<port_no>:AUTO ONCE
Содержание 1145.1010.04/05/06
Страница 10: ......
Страница 20: ......
Страница 22: ......
Страница 24: ...Contents of Chapter 1 R S ZVA 1145 1090 62 I 1 2 E 1 Windows XP Embedded 1 22 Firmware Update 1 22...
Страница 48: ......
Страница 70: ......
Страница 72: ......
Страница 90: ......
Страница 92: ......
Страница 108: ......
Страница 156: ......
Страница 162: ......
Страница 406: ...Display Menu R S ZVA ZVB ZVT 1145 1084 12 4 244 E 6 Stack Tile Horizontally Tile Vertically...
Страница 450: ...Status Reporting System R S ZVA ZVB ZVT 1145 1084 12 5 18 E 1...
Страница 462: ......
Страница 766: ......
Страница 772: ......
Страница 792: ......
Страница 794: ......
Страница 808: ......