R&S
®
ZVA/ZVB/ZVT
MMEMory...
1145.1084.12 6.171
E-6
MMEMory:LOAD:CKIT:SDATa '<conn_name>', '<ckit_name>', MMTHrough |
MFTHrough | FFTHrough | MMLine | MFLine | FFLine | MMATten | MFATten |
FFATten | MMSNetwork | MFSNetwork | FFSNetwork | MOPen | FOPen | MSHort |
FSHort | MOSHort | FOSHort | MREFlect | FREFlect | MMTCh | FMTCh | MSMatch
| FSMatch, '<stdlabel_name>', '<file_name>' [,<port1_no>][,<port2_no>]
Loads cal kit data for a specific connector type, cal kit, and calibration standard from a specified
Touchstone file, assigning a label for the cal data. A restriction on the port assignment may be defined
in addition.
'<conn_name>'
String parameters containing the name of the connector type.
'<ckit_name>'
String parameters containing the name of a calibration kit
available on the analyzer.
Parameters
Standard type; for a description refer to the table of standard
types.
'<stdlabel_name>'
String parameter defining a label for the cal kit data. An empty
string means that no label is defined.
'<file_name>'
String parameter to specify the name and directory of the
Touchstone file to be loaded. A *.s1p file must be used for one-
port standards, a *.s2p file for two-port standards. If no path is
specified the analyzer searches the current directory, to be
queried with
MMEMory:CDIRectory?
.
<port1_no>,
<port2_no>
Port assignment: One port number for one-port standards, two
port number for two-port standards. If the port numbers are
omitted, the cal kit data is valid for all ports.
*RST value
–
SCPI, Command
Types
Device-specific, no query.
Example:
MMEM:LOAD:CKIT:SDAT 'N 50 Ohm','Default
Kit',MOPEN,'Test data','test.s1p',1
Load the file
Test.s1p
from the current
directory in order to define the properties of
an Open (m) standard in the cal kit named
Default Kit
for the
N 50
Ω
connector type
.
Assign the label
Test data
and specify that the
data standard data is only valid for port no. 1.
Connector and cal kit naming conventions
Connector and calibration kit names must be entered as string parameters. The strings contain the
connector and cal kit names used in the
Calibration Kits
dialog; a
Ω
in the name must be replaced by
'Ohm', e.g.:
'NewKit1' denotes the user-defined calibration kit
NewKit1
.
'N 50 Ohm Dummy Kit' denotes the
N 50
Ω
Dummy Kit
.
Содержание 1145.1010.04/05/06
Страница 10: ......
Страница 20: ......
Страница 22: ......
Страница 24: ...Contents of Chapter 1 R S ZVA 1145 1090 62 I 1 2 E 1 Windows XP Embedded 1 22 Firmware Update 1 22...
Страница 48: ......
Страница 70: ......
Страница 72: ......
Страница 90: ......
Страница 92: ......
Страница 108: ......
Страница 156: ......
Страница 162: ......
Страница 406: ...Display Menu R S ZVA ZVB ZVT 1145 1084 12 4 244 E 6 Stack Tile Horizontally Tile Vertically...
Страница 450: ...Status Reporting System R S ZVA ZVB ZVT 1145 1084 12 5 18 E 1...
Страница 462: ......
Страница 766: ......
Страница 772: ......
Страница 792: ......
Страница 794: ......
Страница 808: ......