R&S
®
ZVA/ZVB/ZVT
CALCulate
1145.1084.12 6.65
E-6
CALCulate<Ch/Tr>:LIMit:FAIL?
Returns a 0 or 1, to indicate whether the LIMit test has failed or not.
<Ch/Tr>
Channel number used to identify the active trace.
Response
0 | 1 - 0 represents pass, 1 represents fail.
*RST value
0
SCPI, Command
Types
Confirmed, query only.
Example:
*RST; CALC:LIM:CONT 1 GHZ, 2 GHZ
Define an upper limit line segment in the stimulus range between 1
GHz and 2 GHz, using default response values.
CALC:LIM:STAT ON; CALC:LIM:FAIL?
Switch the limit check on and query the result.
CALCulate<Ch/Tr>:LIMit:LOWer[:DATA]
<numeric_value>,<numeric_value>{,<numeric_value>,<numeric_value>}
Defines the response (y-axis) values of the lower limit line and/or creates new limit line segments.
The commands
CALCulate<Ch/Tr>:LIMit:LOWer[:DATA]
and
CALCulate<Ch/Tr>:LIMit:UPPer[:DATA]
use a fixed numbering scheme for limit line
segments: Upper limit line segments are assigned odd numbers (1, 3, 5,...), lower limit line
segments are assigned even numbers (2, 4, 6,...).
Rules for creating segments
The following rules apply to an active trace with n existing upper and n existing lower limit line
segments:
An odd number of values is rejected; an error message –109,"Missing parameter..." is
generated.
An even number of 2*k values updates or generates k lower limit line segments.
For n > k the response values of all existing lower limit line segments no. 2, 4, ...,2*k are
updated, the existing upper and lower limit line segments no. 2*k+1, ..., 2*n are deleted. The
existing upper limit line segments no. 1, 3, 2*k–1 are not affected.
For n < k the response values of the lower limit line segments no. 2, 4 to 2*n are updated, the
lower limit line segments 2*n+2, 2*n+4,..., 2*k are generated with default stimulus values (see
CALCulate<Ch/Tr>:LIMit:CONTrol[:DATA]
. In addition, the missing upper limit line
segments 2*n+1, 2*n+3,..., 2*k–1 are generated with default stimulus and response values
<Ch/Tr>
Channel number used to identify the active trace.
<numeric_value>
Pair(s) of response values.
Range [def. unit]
Almost no restriction for limit segments; see
Rules for Limit Line Definition.
[dB]
Содержание 1145.1010.04/05/06
Страница 10: ......
Страница 20: ......
Страница 22: ......
Страница 24: ...Contents of Chapter 1 R S ZVA 1145 1090 62 I 1 2 E 1 Windows XP Embedded 1 22 Firmware Update 1 22...
Страница 48: ......
Страница 70: ......
Страница 72: ......
Страница 90: ......
Страница 92: ......
Страница 108: ......
Страница 156: ......
Страница 162: ......
Страница 406: ...Display Menu R S ZVA ZVB ZVT 1145 1084 12 4 244 E 6 Stack Tile Horizontally Tile Vertically...
Страница 450: ...Status Reporting System R S ZVA ZVB ZVT 1145 1084 12 5 18 E 1...
Страница 462: ......
Страница 766: ......
Страница 772: ......
Страница 792: ......
Страница 794: ......
Страница 808: ......