![Rohde & Schwarz 1145.1010.04/05/06 Скачать руководство пользователя страница 757](http://html1.mh-extra.com/html/rohde-and-schwarz/1145-1010-04-05-06/1145-1010-04-05-06_operating-manual_1477759757.webp)
R&S
®
ZVA/ZVB/ZVT
SYSTem...
1145.1084.12 6.295
E-6
SCPI, Command
Types
Confirmed, no query
Example:
INIT:CONT OFF
Activate single sweep mode.
SYST:SETT:UPD ONCE
Update the settings made during the current single sweep
period.
SYSTem:SOUNd:ALARm[:STATe] <Boolean>
Switches alarm sounds on or off.
<Boolean>
OFF - Alarm sounds switched off
ON - Alarm sounds switched on
*RST value
ON
SCPI, Command Types
Device-specific, command or query.
Example:
SYST:SOUN:ALAR OFF; STAT OFF
Switch alarm and status sounds off.
SYSTem:SOUNd:STATus[:STATe] <Boolean>
Switches status sounds on or off.
<Boolean>
OFF - Status sounds switched off
ON - Status sounds switched on
*RST value
ON
SCPI, Command Types
Device-specific, command or query.
Example:
See SYSTem:SOUNd:ALARm[:STATe]
SYSTem:VERSion?
Returns the SCPI version number to which the analyzer complies. The analyzer complies to the final
SCPI version 1999.0
*RST value
–
SCPI, Command Types
Confirmed, query only
Example:
SYST:VERS?
Query the SCPI version. The response is
1999.0
.
Содержание 1145.1010.04/05/06
Страница 10: ......
Страница 20: ......
Страница 22: ......
Страница 24: ...Contents of Chapter 1 R S ZVA 1145 1090 62 I 1 2 E 1 Windows XP Embedded 1 22 Firmware Update 1 22...
Страница 48: ......
Страница 70: ......
Страница 72: ......
Страница 90: ......
Страница 92: ......
Страница 108: ......
Страница 156: ......
Страница 162: ......
Страница 406: ...Display Menu R S ZVA ZVB ZVT 1145 1084 12 4 244 E 6 Stack Tile Horizontally Tile Vertically...
Страница 450: ...Status Reporting System R S ZVA ZVB ZVT 1145 1084 12 5 18 E 1...
Страница 462: ......
Страница 766: ......
Страница 772: ......
Страница 792: ......
Страница 794: ......
Страница 808: ......