R&S
®
ZVA/ZVB/ZVT
SENSe
1145.1084.12 6.205
E-6
The main purpose of the default correction data set is to provide a dummy system error
correction which you can replace with your own, external correction data. You may have acquired the
external data in a previous session or even on an other instrument. If you want to use the external
correction data on the analyzer, simply generate the default data set corresponding to your port
configuration and calibration type and overwrite the default data. For details refer to the program
example below.
<Ch>
Channel number of the calibrated channel.
*RST value
–
SCPI,
Command
Types
Device-specific, no query
Example:
CORR:COLL:METH:DEF 'Test',RSHort,1
Select a one-port normalization at port 1 with a short standard as
calibration type.
CORR:COLL:SAVE:SEL:DEF
Calculate a dummy system error correction for the normalization at port
1. The dummy system error correction provides the reflection tracking
error term 'REFLTRACK'.
CORR:CDAT? 'REFLTRACK',1,0
Query the dummy system error correction term. The response is a 1
(written as 1,0 for the real and imaginary part) for each sweep point (no
attenuation and no phase shift between the analyzer and the calibration
plane).
CORR:CDAT 'REFLTRACK',1,0,<ASCII_data>
Replace the dummy system error correction term with your own
correction data, transferred in ASCII format.
[SENSe<Ch>:]CORRection:CONNection '<conn_name>', TEM | WGUide,
GENDer | NGENder, <perm_rel>, <imped>
Configures the user-defined connector types.
<Ch>
Channel number
'<conn_name>'
Name of the user-defined connectors, string parameter
TEM | WGUide
Transverse electric or waveguide type propagation mode
GENDer | NGENder
Polar connector type (m/f)
Sexless connector type
<perm_rel>
Relative permittivity
Range [def. unit]
0.0000000001 to 1000. [–] UP/DOWN/MIN/MAX parameters are
not available for this command
*RST value
–
<imped>
For TEM type connectors: reference impedance in
Ω
(without
unit),
For WGUide type connectors: cutoff frequency in Hz (without
unit)
Содержание 1145.1010.04/05/06
Страница 10: ......
Страница 20: ......
Страница 22: ......
Страница 24: ...Contents of Chapter 1 R S ZVA 1145 1090 62 I 1 2 E 1 Windows XP Embedded 1 22 Firmware Update 1 22...
Страница 48: ......
Страница 70: ......
Страница 72: ......
Страница 90: ......
Страница 92: ......
Страница 108: ......
Страница 156: ......
Страница 162: ......
Страница 406: ...Display Menu R S ZVA ZVB ZVT 1145 1084 12 4 244 E 6 Stack Tile Horizontally Tile Vertically...
Страница 450: ...Status Reporting System R S ZVA ZVB ZVT 1145 1084 12 5 18 E 1...
Страница 462: ......
Страница 766: ......
Страница 772: ......
Страница 792: ......
Страница 794: ......
Страница 808: ......