
R&S
®
ZVA/ZVB/ZVT
Measured Quantities
1145.1084.12 3.27
E-1
•
S
22
is the output reflection coefficient, defined as the ratio of the wave quantities b
2
(reverse
measurement with matched input, b
2,rev
in the figure above and a
1
= 0) to a
2
, measured at
PORT 2.
Meaning of squared amplitudes
The squared amplitudes of the incident and outgoing waves and of the matrix elements have a simple
meaning:
|a1|
2
Available incident power
at the input of a two-port (= the power provided by a
generator with a source impedance equal to the reference impedance Z
0
)
|a2|
2
Available incident power
at the output
|b1|
2
Reflected power at the input of a two-port
|b2|
2
Reflected power at the output
10*log|S
11
|
2
(= 20*log|S
11
|)
Reflection loss at the input
10*log|S
22
|
2
Reflection loss at the output
10*log|S
21
|
2
Insertion loss at the input
10*log|S
12
|
2
Insertion loss at the output
Multiport Parameters
The multiport S-parameters extend the standard 2-port S-parameters to a larger number of incoming
and outgoing waves. For a 4-port DUT,
∗
=
a
a
a
a
S
S
S
S
S
S
S
S
S
S
S
S
S
S
S
S
b
b
b
b
4
3
2
1
44
43
42
41
34
33
32
31
24
23
22
21
14
13
12
11
4
3
2
1
,
where again a
i
(i = 1 to 4) denote the incident, b
i
(i = 1 to 4) denote the outgoing waves, and the S-
parameters are expressed as S
<out>< in>
.
The indices of the S-parameters described so far number the output and input ports of a DUT; the
parameters are referred to as single-ended S-parameters. The S-parameter description can also be
used to differentiate between different propagation modes of the waves at the output and input ports.
This results in the so-called mixed mode S-parameters. The analyzer measures either single-ended or
mixed mode S-parameters.
Impedance Parameters
An impedance is the complex ratio between a voltage and a current. The analyzer provides two
independent sets of impedance parameters, essentially based on different n-port circuit models:
•
Matched-circuit impedances, converted from reflection S-parameters
•
Open-circuit Z-parameters
Converted Impedances
The converted, matched-circuit impedances describe the impedances of a DUT that is terminated at its
outputs with the reference impedance Z
0i
displayed in the
Port Configuration
dialog. i numbers the
analyzer/DUT port.
Содержание 1145.1010.04/05/06
Страница 10: ......
Страница 20: ......
Страница 22: ......
Страница 24: ...Contents of Chapter 1 R S ZVA 1145 1090 62 I 1 2 E 1 Windows XP Embedded 1 22 Firmware Update 1 22...
Страница 48: ......
Страница 70: ......
Страница 72: ......
Страница 90: ......
Страница 92: ......
Страница 108: ......
Страница 156: ......
Страница 162: ......
Страница 406: ...Display Menu R S ZVA ZVB ZVT 1145 1084 12 4 244 E 6 Stack Tile Horizontally Tile Vertically...
Страница 450: ...Status Reporting System R S ZVA ZVB ZVT 1145 1084 12 5 18 E 1...
Страница 462: ......
Страница 766: ......
Страница 772: ......
Страница 792: ......
Страница 794: ......
Страница 808: ......