Channel Menu
R&S
®
ZVA/ZVB/ZVT
1145.1084.12 4.162
E-6
Measurements do not necessarily require all of the physical and logical ports of the network analyzer.
To save measurement time, it is recommended to define the ports that are actually used in the current
test setup
(Define Measured Ports)
. Unused ports will not be considered for the calculation of mixed
mode, Z- and Y-parameters.
In the four tabs of the
Balanced and Measured Ports
dialog it is possible to
•
Select one of the predefined balanced port configurations
(Predefined Configs).
•
Define a logical port based on two physical ports and independent reference impedances for
the calculation of common mode and differential mode parameters
(Def. Balanced Port)
.
•
Dissolve a single logical port so that the two involved ports are again used as physical ports
(Dissolve Bal. Port)
.
•
Select the ports that can be used for the measurement
(Define Measured Ports).
Predefined Configs
The
Predefined Configs
tab of the
Balanced and Measured Ports
dialog provides the most commonly
used balanced port configurations for the analyzer.
Содержание 1145.1010.04/05/06
Страница 10: ......
Страница 20: ......
Страница 22: ......
Страница 24: ...Contents of Chapter 1 R S ZVA 1145 1090 62 I 1 2 E 1 Windows XP Embedded 1 22 Firmware Update 1 22...
Страница 48: ......
Страница 70: ......
Страница 72: ......
Страница 90: ......
Страница 92: ......
Страница 108: ......
Страница 156: ......
Страница 162: ......
Страница 406: ...Display Menu R S ZVA ZVB ZVT 1145 1084 12 4 244 E 6 Stack Tile Horizontally Tile Vertically...
Страница 450: ...Status Reporting System R S ZVA ZVB ZVT 1145 1084 12 5 18 E 1...
Страница 462: ......
Страница 766: ......
Страница 772: ......
Страница 792: ......
Страница 794: ......
Страница 808: ......