Channel
Menu
1145.1084.12 4.133
E-6
Sweep
The
Sweep
submenu defines the scope of measurement in the current channel. This includes the
sweep type with various parameters, the trigger conditions and the periodicity of the measurement.
Sweeps
A sweep is a series of consecutive measurements taken over a specified sequence of stimulus values.
It represents the basic measurement cycle of the analyzer.
The analyzer can perform sweeps at constant power but variable frequency (frequency sweeps),
sweeps at constant frequency but variable power (power sweeps), and sweeps at constant power and
frequency that are repeated in time (Time/CW Trigger Event sweeps); see Sweep Type.
The sweeps are further specified by the number of measurement points, the total measurement time
and the trigger mode. A measurement may consist of a single sweep or a series of sweeps repeated
continuously.
On the other hand, depending on the measurement task and the measured quantity, the measurement
at each point can consist of several partial measurements with definite hardware settings.
The
Sweep
menu provides the following settings:
•
Sweep Type
defines the sweep variable (frequency/power/time) and the position of sweep
points in the sweep range.
•
Number of Points
sets the total number of measurement points over the sweep range.
•
Frequency Step Size
sets the distance between two consecutive frequency sweep points.
•
Sweep Time
varies the measurement time for a sweep or delays the start of each sweep.
•
Trigger
selects the trigger mode for the measurement.
•
Manual Trigger
while the
Manual
trigger mode is active generates a trigger event. This softkey
is also available in the
Trigger
submenu.
•
If
Single (All Chans)
is selected the measurement is terminated after a single sweep or a group
of single sweeps defined in the
Define Restart
dialog.
•
Restart
aborts the current measurement and restarts a new sweep sequence.
•
Define Restart
opens a dialog to specify how many sweeps are repeated.
Содержание 1145.1010.04/05/06
Страница 10: ......
Страница 20: ......
Страница 22: ......
Страница 24: ...Contents of Chapter 1 R S ZVA 1145 1090 62 I 1 2 E 1 Windows XP Embedded 1 22 Firmware Update 1 22...
Страница 48: ......
Страница 70: ......
Страница 72: ......
Страница 90: ......
Страница 92: ......
Страница 108: ......
Страница 156: ......
Страница 162: ......
Страница 406: ...Display Menu R S ZVA ZVB ZVT 1145 1084 12 4 244 E 6 Stack Tile Horizontally Tile Vertically...
Страница 450: ...Status Reporting System R S ZVA ZVB ZVT 1145 1084 12 5 18 E 1...
Страница 462: ......
Страница 766: ......
Страница 772: ......
Страница 792: ......
Страница 794: ......
Страница 808: ......