CALCulate
R&S
®
ZVA/ZVB/ZVT
1145.1084.12 6.122
E-6
SCPI, Command
Types
Device-specific, command or query.
Example:
See
CALCulate<Ch>:TRANsform:VNETworks:BALanced:DEEMbedding[:STATe]
CALCulate<Ch>:TRANsform:VNETworks:BALanced:EMBedding<Log_pt>:PARa
meters:C<no> STSC | SCST | CSSL | LSSC | CSSC | SLCS | SCLS |
SCCS,<numeric_value>
Specifies the capacitance values C1, C2, C3 in the different circuit models for balanced port
embedding.
<Ch>
Channel number
<Log_pt>
Logical port number
<no>
Number of capacitance in circuit model. If unspecified the numeric
suffix is set to 1.
Range [def. unit]
1 to 3 [–] The number of capacitances depends on the selected
circuit model.
Parameters
Possible circuit models (character data); see circuit models.
*RST value
– (see command type description below)
<numeric_value>
Capacitance C<no> for the specified circuit model.
Range [def. unit]
–1mF to 1 mF [F]. The increment is 1 fF (10
–15
F).
*RST value
1 pF (10
–12
F)
SCPI, Command
Types
Device-specific, command or query. In the query form, the
<numeric_value> parameter must be omitted. The command
returns the capacitance value for the specified circuit model.
Example:
*RST; CALC:TRAN:VNET:BAL:EMB:PAR:C2? CSSL
Query the default capacitance C2 for the
Serial Cs, shunt L
circuit
model. The response is 1E-012 (1 pF).
CALC:TRAN:VNET:BAL:EMB:PAR:C2 CSSL, 2.2E-12
Increase the capacitance to 2.2 pF.
CALCulate<Ch>:TRANsform:VNETworks:BALanced:EMBedding<Log_pt>:PARa
meters:L<no> STSL | SLST | CSSL | LSSC | LSSL | SLCS | SCLS |
SLLS,<numeric_value>
Specifies the inductance values L1, L2, L3 in the different circuit models for balanced port embedding.
<Ch>
Channel number
<Log_pt>
Logical port number
<no>
Number of inductance in circuit model. If unspecified the numeric
suffix is set to 1.
Range [def. unit]
1 to 3 [–] The number of inductances depends on the selected
circuit model.
Parameters
Possible circuit models (character data); see circuit models.
Содержание 1145.1010.04/05/06
Страница 10: ......
Страница 20: ......
Страница 22: ......
Страница 24: ...Contents of Chapter 1 R S ZVA 1145 1090 62 I 1 2 E 1 Windows XP Embedded 1 22 Firmware Update 1 22...
Страница 48: ......
Страница 70: ......
Страница 72: ......
Страница 90: ......
Страница 92: ......
Страница 108: ......
Страница 156: ......
Страница 162: ......
Страница 406: ...Display Menu R S ZVA ZVB ZVT 1145 1084 12 4 244 E 6 Stack Tile Horizontally Tile Vertically...
Страница 450: ...Status Reporting System R S ZVA ZVB ZVT 1145 1084 12 5 18 E 1...
Страница 462: ......
Страница 766: ......
Страница 772: ......
Страница 792: ......
Страница 794: ......
Страница 808: ......