Program Examples
R&S
®
ZVA/ZVB/ZVT
1145.1084.12 7.2
E-1
//
// Select active traces for the new created channels 1 - 4
:CALCulate1:PARameter:SELect 'Trc1'
:CALCulate2:PARameter:SELect 'Trc2'
:CALCulate3:PARameter:SELect 'Trc3'
:CALCulate4:PARameter:SELect 'Trc4'
//
// Create diagrams 1 - 4
:DISPlay:WINDow1:STATe ON
:DISPlay:WINDow2:STATe ON
:DISPlay:WINDow3:STATe ON
:DISPlay:WINDow4:STATe ON
//
// Assign trace 'Trc1' - 'Trc4' to a window = diagram area 1 - 4
:DISPlay:WINDow1:TRACe1:FEED 'Trc1'
:DISPlay:WINDow2:TRACe1:FEED 'Trc2'
:DISPlay:WINDow3:TRACe1:FEED 'Trc3'
:DISPlay:WINDow4:TRACe1:FEED 'Trc4'
//
// Query trace names assigned to window trace
:DISPlay:WINDow1:TRACe1:FEED?
:DISPlay:WINDow2:TRACe1:FEED?
:DISPlay:WINDow3:TRACe1:FEED?
:DISPlay:WINDow4:TRACe1:FEED?
Calibration
The following example shows you how to perform a multiport calibration of the analyzer.
//
Reset
the analyzer
*RST
//
//
Set cal kit
as active kit for N50
:SENSE:CORRECTION:CKIT:N50:SELECT 'N 50 Ohm Dummy Kit'
//
//
Select connectors
for the ports
:SENSE1:CORRECTION:COLLECT:CONNECTION1 N50MALE
:SENSE1:CORRECTION:COLLECT:CONNECTION2 N50MALE
//
Содержание 1145.1010.04/05/06
Страница 10: ......
Страница 20: ......
Страница 22: ......
Страница 24: ...Contents of Chapter 1 R S ZVA 1145 1090 62 I 1 2 E 1 Windows XP Embedded 1 22 Firmware Update 1 22...
Страница 48: ......
Страница 70: ......
Страница 72: ......
Страница 90: ......
Страница 92: ......
Страница 108: ......
Страница 156: ......
Страница 162: ......
Страница 406: ...Display Menu R S ZVA ZVB ZVT 1145 1084 12 4 244 E 6 Stack Tile Horizontally Tile Vertically...
Страница 450: ...Status Reporting System R S ZVA ZVB ZVT 1145 1084 12 5 18 E 1...
Страница 462: ......
Страница 766: ......
Страница 772: ......
Страница 792: ......
Страница 794: ......
Страница 808: ......