![Rohde & Schwarz 1145.1010.04/05/06 Скачать руководство пользователя страница 249](http://html1.mh-extra.com/html/rohde-and-schwarz/1145-1010-04-05-06/1145-1010-04-05-06_operating-manual_1477759249.webp)
R&S
®
ZVA/ZVB/ZVT
Trace
Menu
1145.1084.12 4.87
E-6
Markers are available for all diagram types (Trace – Format).
Marker 1/2/3
Creates the markers numbered 1, 2, and 3, respectively, and assigns them to the active trace (toggle
function).
Marker 1/2/3
opens the numeric entry bar to define the marker position
(Stimulus Mkr 1/2/3)
.
The default position is the center of the sweep range.
On closing the
Stimulus Mkr 1/2/3
numeric entry bar a marker symbol (triangle) labeled
Mkr <n>
is
positioned on the trace and the marker coordinates are displayed in the
Info Field
.
Activating and moving markers
To select one of several markers as an active marker, do one of the following:
•
Click the marker symbol.
•
Click the marker line in the marker info field.
To change the position of the active marker on the trace use one of the following methods:
•
Drag-and-drop the marker symbol to the desired position.
•
Click the
Marker <n>
or
Ref. Marker
softkey to call up the entry bar for the new stimulus value.
•
Right-click the diagram area or select
Mkr. Properties
to call up the
Marker Properties
dialog
and select the new stimulus value.
•
Use the
Search
functions to place the marker to a specific point on the trace.
If the marker position is defined explicitly by entering a numeric value, the marker position can be
outside the sweep range. If it is just varied using the rollkey, the mouse or the cursor keys, it
always remains within the sweep range. If the position of a marker outside the sweep range is
varied, it is automatically moved to the start or stop value of the sweep range, whichever is
closer.
Remote control:
CALCulate<Ch/Tr>:MARKer<Mk>[:STATe] ON
CALCulate<Ch/Tr>:MARKer<Mk>:Y?
Ref. Marker
Creates a reference marker and assigns it to the active trace (toggle function).
Ref. Marker
opens the
numeric entry bar to define the marker position
(Stimulus Ref Mkr)
. The default position is the center of
the sweep range.
On closing the
Stimulus Ref Mkr
numeric entry bar a marker symbol (triangle) labeled
Ref
is positioned
on the trace and a line indicating
Ref
plus the marker coordinates is inserted in the marker
info field
.
The reference marker defines the reference value for all markers that are in
Delta Mode.
Remote control:
CALCulate<Ch/Tr>:MARKer<Mk>:REFerence[:STATe] ON
CALCulate<Ch/Tr>:MARKer<Mk>:REFerence:Y?
Содержание 1145.1010.04/05/06
Страница 10: ......
Страница 20: ......
Страница 22: ......
Страница 24: ...Contents of Chapter 1 R S ZVA 1145 1090 62 I 1 2 E 1 Windows XP Embedded 1 22 Firmware Update 1 22...
Страница 48: ......
Страница 70: ......
Страница 72: ......
Страница 90: ......
Страница 92: ......
Страница 108: ......
Страница 156: ......
Страница 162: ......
Страница 406: ...Display Menu R S ZVA ZVB ZVT 1145 1084 12 4 244 E 6 Stack Tile Horizontally Tile Vertically...
Страница 450: ...Status Reporting System R S ZVA ZVB ZVT 1145 1084 12 5 18 E 1...
Страница 462: ......
Страница 766: ......
Страница 772: ......
Страница 792: ......
Страница 794: ......
Страница 808: ......