Trace Menu
R&S
®
ZVA/ZVB/ZVT
1145.1084.12 4.96
E-6
Target Search >
Activates the search to the right of the active marker position and sets the active marker to the defined
target value. The target search range is between the active marker position and the end (Stop) of the
sweep range. If the active trace contains no markers, a marker
Mkr 1
is created to indicate the search
result, and the target search range starts at the beginning (Start) of the sweep range.
If the target value occurs at several stimulus values, the marker is placed to the search result with the
smallest stimulus value. The other measurement points with the same target value can be located using
Target Search >
repeatedly.
If the target is not found (e.g. because the active trace doesn't contain the target value), then the active
marker is not moved away from its original position.
Remote control:
CALCulate<Ch/Tr>:MARKer<Mk>:FUNCtion:EXECute RTARget
CALCulate<Ch/Tr>:MARKer<Mk>:FUNCtion:RESult?
< Target Search
Activates the search to the left of the active marker position and sets the active marker to the defined
target value. The target search range is between the (Start) of the sweep range and the active marker
position. If the active trace contains no markers, a marker
Mkr 1
is created to indicate the search result,
and the target search range starts at the end (Stop) of the sweep range.
If the target value occurs at several stimulus values, the marker is placed to the search result with the
smallest stimulus value. The other measurement points with the same target value can be located using
Target Search >
repeatedly.
If the target is not found (e.g. because the active trace doesn't contain the target value), then the active
marker is not moved away from its original position.
Remote control:
CALCulate<Ch/Tr>:MARKer<Mk>:FUNCtion:EXECute LTARget
CALCulate<Ch/Tr>:MARKer<Mk>:FUNCtion:RESult?
Target Value
Opens an dialog to select a marker for the target search and define the target value.
It is possible to define up to ten different target values for each trace and assign them to the markers
no. 1 to 10. The input fields in the
Target Search
dialog are used to select the markers and define the
associated search ranges:
•
Marker
selects one of the ten markers that can be assigned to the trace. If a selected marker
does not exist, it is created as soon as
On
is checked. A created marker is displayed in the
center of the search range.
•
Value
selects the target value to be assigned to the selected marker. The target
Value
is
entered with the unit of the active trace.
Remote control:
CALCulate<Ch/Tr>:MARKer<Mk>:TARget
Содержание 1145.1010.04/05/06
Страница 10: ......
Страница 20: ......
Страница 22: ......
Страница 24: ...Contents of Chapter 1 R S ZVA 1145 1090 62 I 1 2 E 1 Windows XP Embedded 1 22 Firmware Update 1 22...
Страница 48: ......
Страница 70: ......
Страница 72: ......
Страница 90: ......
Страница 92: ......
Страница 108: ......
Страница 156: ......
Страница 162: ......
Страница 406: ...Display Menu R S ZVA ZVB ZVT 1145 1084 12 4 244 E 6 Stack Tile Horizontally Tile Vertically...
Страница 450: ...Status Reporting System R S ZVA ZVB ZVT 1145 1084 12 5 18 E 1...
Страница 462: ......
Страница 766: ......
Страница 772: ......
Страница 792: ......
Страница 794: ......
Страница 808: ......