[SENSe<Ch>:]CORRection:COLLect[:ACQuire]:RSAVe <Boolean>
Activates or deactivates the calibration mode where the raw measurement data of the standards is
stored after the calibration is completed. The setting is valid for the current calibration, where it
overwrites the global setting
(
[SENSe<Ch>:]CORRection:COLLect[:ACQuire]:RSAVe:DEFault
). A new calibration deletes
the calibration data acquired in previous calibrations.
<Ch>
Channel number of the calibrated channel.
<Boolean>
ON | OFF - Keep measurement data on or off.
*RST value
OFF
SCPI, Command
Types
Device-specific, command or query
Example:
CORR:COLL:RSAV:DEF ON
Generally keep the raw measurement data of the standards after
the calibration is completed.
CORR:COLL:METH:DEF 'Test',RSHort,1
Select a one-port normalization at port 1 with a short standard as
calibration type.
CORR:COLL:SEL SHOR,1
Measure a short standard connected to port 1 and store the raw
measurement results of this standard.
CORR:COLL:RSAV OFF
To save disk space, delete the current raw calibration data after
the calibration is completed.
CORR:COLL:SAVE:SEL
Calculate the system error correction data and apply them to the
active channel.
[SENSe<Ch>:]CORRection:COLLect[:ACQuire]:RSAVe:DEFault <Boolean>
Activates or deactivates the calibration mode where the raw measurement data of the standards is
stored after the calibration is completed. The setting remains valid for all subsequent calibrations until it
is changed explicitly. A new calibration deletes the calibration data acquired in previous calibrations.
<Ch>
Channel number of the calibrated channel.
<Boolean>
ON | OFF - Keep measurement data on or off.
*RST value
OFF
SCPI, Command Types
Device-specific, command or query
Example:
See
[SENSe<Ch>:]
CORRection:COLLect[:AQUire]:RSAVe
.
[SENSe<Ch>:]CORRection:COLLect[:ACQuire]:SELected THRough | OPEN |
SHORt | MATCh | NET | ATT | REFL | SLIDE | LINE,<port_no>[,<port_no>]
Starts a calibration measurement in order to acquire measurement data for the selected standards. The
standards are reflection or transmission standards and can be connected to arbitrary analyzer ports.
Содержание 1145.1010.04/05/06
Страница 10: ......
Страница 20: ......
Страница 22: ......
Страница 24: ...Contents of Chapter 1 R S ZVA 1145 1090 62 I 1 2 E 1 Windows XP Embedded 1 22 Firmware Update 1 22...
Страница 48: ......
Страница 70: ......
Страница 72: ......
Страница 90: ......
Страница 92: ......
Страница 108: ......
Страница 156: ......
Страница 162: ......
Страница 406: ...Display Menu R S ZVA ZVB ZVT 1145 1084 12 4 244 E 6 Stack Tile Horizontally Tile Vertically...
Страница 450: ...Status Reporting System R S ZVA ZVB ZVT 1145 1084 12 5 18 E 1...
Страница 462: ......
Страница 766: ......
Страница 772: ......
Страница 792: ......
Страница 794: ......
Страница 808: ......